D-12
CO2 Hardware
Error
H
Errors occurred to:
1. External A/D sampling 2.5 V
2. 12V power supply voltage
3. Internal A/D sampling 2.5 V
4. Pump.
5. 3-way valve.
CO2 Sampleline
Occluded
M An error or occlusion occurred to the sampling line.
CO2 Zero Failed
H
Deviation of gain input signal is too great to be adjusted.
Namely, it cannot be adjusted within the normal range: 3.5
V±100 mV.
CO2 Cal. Failed
M The difference between the measured standard gas
concentration and the specified standard gas concentration
exceeded 40% of the specified standard gas concentration.
Or, an illegal calibration parameter was obtained. The
normal parameter calibration range is within 0.2 to 2.5.
CO2 System Error
H
Multiple system errors occurred.
CO2 No Watertrap
M The CO
2
watertrap fell off or was not connected.
EtCO2 Overrange
H
FiCO2 Overrange
H
The measured value is outside of the measurement range.
Contact your service personnel.
CO2 Check Airway M Airway error.
CO2 No
Sampleline
L
Make sure if the sampling line is already connected.
CO2 Main Board
Error
H
CO2 Check Sensor
or Main Board
M
CO2 Replace
Scrubber&Pump
M
CO2 Replace
Sensor
M
CO2 15V
Overrange
M
CO2 module malfunction. Re-plug the module or restart
the anesthesia machine.
BIS Init Error
H
BIS Selftest Error
H
BIS Comm Stop
H
BIS Comm Error
H
Module malfunction or communication failure between
the module and main unit. Re-plug the module, restart the
machine, or plug the module onto other main unit.
BIS
module
BIS ALM LMT
ERR
M The alarm limits are changes accidentally. Contact your
service personnel.
Содержание WATO EX-65
Страница 1: ...WATO EX 65 Anesthesia Machine Operator s Manual ...
Страница 2: ......
Страница 8: ...VI FOR YOUR NOTES ...
Страница 16: ...8 FOR YOUR NOTES ...
Страница 25: ...2 3 2 2 Equipment Appearance 2 2 1 Front View Display and control panel ...
Страница 27: ...2 5 Breathing system ...
Страница 29: ...2 7 2 2 2 Rear View Power supply ...
Страница 31: ...2 9 Anesthesia information system CIS ...
Страница 42: ...3 8 FOR YOUR NOTES ...
Страница 66: ...4 24 FOR YOUR NOTES ...
Страница 108: ...7 10 7 8 AGSS Transfer Tube Maintenance Check the tube of the AGSS transfer system Replace it if it is damaged ...
Страница 148: ...11 8 FOR YOUR NOTES ...
Страница 152: ...12 4 FOR YOUR NOTES ...
Страница 154: ...13 2 13 1 1 Breathing System Diagrams 14 15 16 17 18 19 20 21 22 1 2 3 4 5 6 7 8 9 10 11 12 13 ...
Страница 168: ...13 16 5 Turn the rotary handle clockwise for 90 degrees 6 Let the rotary handle fall to lock the sodalime canister ...
Страница 222: ...A 6 FOR YOUR NOTES ...
Страница 244: ...B 22 FOR YOUR NOTES ...
Страница 250: ...C 6 FOR YOUR NOTES ...
Страница 264: ...D 14 FOR YOUR NOTES ...
Страница 274: ......
Страница 275: ......
Страница 276: ...P N 046 000203 00 1 0 ...