5-16
The circuit adapter test fixture can be mounted either onto the circuit adapter or onto the
removed patient circuit. The following pictures show the test fixture mounted in position.
If it is hard to plug in and out the test fixture, apply a layer of KRYTOX lubricant (BOM
number: M6F-020003---) to the seals (as shown below).
Seal
(
M6M-010058---
)
Seal
(
M6M-010063---
)
Seal
(
M6M-010006---
)