6-31
3. Unscrew the four screws as shown below to remove the cover plate of the isolation
transformer mounting box.
4. Disconnect the power cables from the auxiliary electrical outlet and from the isolation
transformer to take out the isolation transformer assembly.
5. Unscrew the four screws fixing the handle to remove the handle, as shown below.
Disconnect the power cable from
the auxiliary electrical outlet
Disconnect the power cable from
the isolation transformer