5-23
Note 1: The Nos. of the PU tubes described here correspond to the gas tube Nos. when O2,
N2O and AIR are all configured. In case of other gas supply configurations, if the PU tube
is numbered differently, notes will be given in brackets. For Nos. of all PU tubes, refer
to
6.4.2Pneumatic Connections.
II N2O supply inlet assembly:
1. Turn off the pipeline gas supply. Pull out No.48 PU tube which connects the N2O
supply inlet assembly to other assembly. The end of the tube which connects the N2O
supply inlet assembly is not pulled out but the other end is.
2. Connect the pulled-out tube end to the test pressure gauge through 3106-04-06 adapter
connector, as shown below.
3. Turn on N2O and O2 pipeline supplies. Adjust the regulator of the N2O supply inlet
assembly to cause the reading on the test pressure gauge to be same to that acquired in
step 4 of “I O2 supply inlet assembly” (for operations of the regulator, refer to section
5.3.2.4Adjust the Regulator of the Pipeline Gas Supply Inlet Assembly
). Record the
reading on the N2O pipeline pressure gauge.
4. Turn off N2O pipeline supply and bleed the residual pressure by turning on the N2O
flow regulator.
5. Insert the pulled-out tube properly.
6. Pull out No.40 PU tube which connects the N2O supply inlet assembly to the N2O
pipeline pressure gauge. The end of the tube which connects the N2O supply inlet
assembly is not pulled out but the other end is.
7. Connect the pulled-out tube end to the test pressure gauge, as shown below.
Содержание WATO EX-55
Страница 1: ...WATO EX 55 65 Anesthesia Machine Service Manual ...
Страница 2: ......
Страница 4: ...II FOR YOUR NOTES ...
Страница 6: ...IV FOR YOUR NOTES ...
Страница 12: ...6 FOR YOUR NOTES ...
Страница 18: ...2 2 2 2 Gas Flow 2 2 1 Pneumatic Circuit Diagram vaporizer vaporizer Patient Waste gas disposal Patient Air ...
Страница 58: ...3 22 FOR YOUR NOTES ...
Страница 133: ...5 13 Negative pressure ball Circuit adapter test fixture Flow sensor pressure sampling pipeline test fixture ...
Страница 134: ...5 14 Vaporizer manifold test fixture Anesthesia machine calibration device 1 MPa 10bar test pressure gauge ...
Страница 242: ...6 54 6 4 1 Electrical Connections Electrical Circuit Diagram ...
Страница 260: ...6 72 FOR YOUR NOTES ...
Страница 261: ...7 1 7 Parts 7 1 Main Unit 7 1 1 Main Unit 7 1 1 1 Exploded View ...
Страница 321: ......
Страница 322: ...P N 0621 20 78654 1 1 ...