6-2
6.1 Prepare for Disassembly
6.1.1 Tools
During parts disassembling and replacing, the following tools may be required:
Metric Allen wrench (2.5#, 3#, 4#, 5#, 8#)
Phillips screwdriver
Diagonal pliers
Flathead screwdriver
Metric M3 and M4 socket screwdriver
Flowmeter calibration fixture
Adjustable wrench
Tweezers
M16 nut mounting fixture (0621-J26-1)
6.1.2 Preparations
Before disassembly,
Make sure that the anesthesia machine is turned off and disconnected from the AC
power source.
Bleed the gas pressure inside the anesthesia machine as described below.
Disconnect all pipeline and cylinder gas supplies.
Prepare the tools required for disassembly.
Maneuver the anesthesia machine to an appropriate location and then step down the
four caster brakes to fix the machine.
CAUTION
z
The internal parts may be contaminated during long-term use of the equipment.
Wear special gloves during disassembling and inspecting.
Содержание WATO EX-55
Страница 1: ...WATO EX 55 65 Anesthesia Machine Service Manual ...
Страница 2: ......
Страница 4: ...II FOR YOUR NOTES ...
Страница 6: ...IV FOR YOUR NOTES ...
Страница 12: ...6 FOR YOUR NOTES ...
Страница 18: ...2 2 2 2 Gas Flow 2 2 1 Pneumatic Circuit Diagram vaporizer vaporizer Patient Waste gas disposal Patient Air ...
Страница 58: ...3 22 FOR YOUR NOTES ...
Страница 133: ...5 13 Negative pressure ball Circuit adapter test fixture Flow sensor pressure sampling pipeline test fixture ...
Страница 134: ...5 14 Vaporizer manifold test fixture Anesthesia machine calibration device 1 MPa 10bar test pressure gauge ...
Страница 242: ...6 54 6 4 1 Electrical Connections Electrical Circuit Diagram ...
Страница 260: ...6 72 FOR YOUR NOTES ...
Страница 261: ...7 1 7 Parts 7 1 Main Unit 7 1 1 Main Unit 7 1 1 1 Exploded View ...
Страница 321: ......
Страница 322: ...P N 0621 20 78654 1 1 ...