5-56
(
3
)
Occlude the pulled-out tube ends by using two 3106-10-00 adapter connectors
and two 3126-10-00 tube plugs.
(
4
)
Repeat steps 3 through 7 in “4 Leak test of all pipelines on the circuit adapter”. If
the test is failed, it indicates that the connectors of the circuit adapter or seals are
damaged. If there is no leak, insert the pulled-out tubes into the ACGO assembly. Note
that the relative position of the black wires on the No.21 and 22 yellow PU tubes
should be same to that before pulled out.
(
5
)
Pull out No.52 and 53 PU tubes which connect the O2 flush button assembly and
the vaporizer manifold assembly to the ACGO assembly. The ends of the tubes which
connect the ACGO assembly are pulled out but the other ends are not.
(
6
)
Occlude the pulled-out tube ends by using 3126-06-00 and 3126-08-00 tube plugs,
as shown below.
(
7
)
Repeat steps 3 through 7 in “4 Leak test of all pipelines on the circuit adapter”. If
the test is failed, it indicates the ACGO assembly is damaged. Check the seals in the
ACGO assembly and replace the damaged seals.
Содержание WATO EX-55
Страница 1: ...WATO EX 55 65 Anesthesia Machine Service Manual ...
Страница 2: ......
Страница 4: ...II FOR YOUR NOTES ...
Страница 6: ...IV FOR YOUR NOTES ...
Страница 12: ...6 FOR YOUR NOTES ...
Страница 18: ...2 2 2 2 Gas Flow 2 2 1 Pneumatic Circuit Diagram vaporizer vaporizer Patient Waste gas disposal Patient Air ...
Страница 58: ...3 22 FOR YOUR NOTES ...
Страница 133: ...5 13 Negative pressure ball Circuit adapter test fixture Flow sensor pressure sampling pipeline test fixture ...
Страница 134: ...5 14 Vaporizer manifold test fixture Anesthesia machine calibration device 1 MPa 10bar test pressure gauge ...
Страница 242: ...6 54 6 4 1 Electrical Connections Electrical Circuit Diagram ...
Страница 260: ...6 72 FOR YOUR NOTES ...
Страница 261: ...7 1 7 Parts 7 1 Main Unit 7 1 1 Main Unit 7 1 1 1 Exploded View ...
Страница 321: ......
Страница 322: ...P N 0621 20 78654 1 1 ...