5-10
Flowmeter Cal.
Data Error 01
H Uncalibrated
data
were
empty. Or, the number of
calibrated electronic
flowmeters was less than
that of tubes configured.
1. Confirm and change the
electronic flowmeter until it is
correctly configured.
2. Calibrate the electronic
flowmeter again.
Flowmeter Cal.
Data Error 02
H
Calibration data error
Calibrate the electronic flowmeter
again.
Flowmeter
Comm Error
H
Flowmeter
Comm Stop
H
The electronic flowmeter
failed to communicate with
the main system normally.
1. Replace the communication
cable for the electronic flowmeter.
2. Replace the electronic flowmeter
board.
N2O Flow Too
High
L
AIR Flow Too
High
L
O2 Flow Too
High
L
Gas flow exceeded 10
L/min.
1. Turn off other gas flow. Compare
the concerned gas flow with the
measurement result displayed on
the glass tube flowmter.
2. Use test tools. Turn on ACGO.
Measure fresh gas flow at the
inspiration connector. Measure the
measurement error of electronic
flowmeter.
3. If the error exceeds the standard,
calibrate the electronic flowmeter
again.
Flowmeter Zero
Failed
L
Zero point error or
three-way valve failure
1. Switch off fresh gas delivery.
Perform zeroing.
2. If the problem persists, check the
three-way valve component of the
electronic flowmeter module,
control circuit and cable.
3. Replace the three-way valve
assembly (throttling device).
4. Replace the electronic flowmeter
board.
Notes:
1. If special calibration equipment is not available, return the electronic flowmeter to the
factory for calibration together with the throttling device of the machine.
2. If the electronic flowmeter board is returned to the factory for replacement, the
throttling device of the machine must also be returned for replacement.
3. If special calibration equipment is available, you can just replace the electronic
flowmeter board. It can be used normally only after it is calibrated on the machine.
Содержание WATO EX-55
Страница 1: ...WATO EX 55 65 Anesthesia Machine Service Manual ...
Страница 2: ......
Страница 4: ...II FOR YOUR NOTES ...
Страница 6: ...IV FOR YOUR NOTES ...
Страница 12: ...6 FOR YOUR NOTES ...
Страница 18: ...2 2 2 2 Gas Flow 2 2 1 Pneumatic Circuit Diagram vaporizer vaporizer Patient Waste gas disposal Patient Air ...
Страница 58: ...3 22 FOR YOUR NOTES ...
Страница 133: ...5 13 Negative pressure ball Circuit adapter test fixture Flow sensor pressure sampling pipeline test fixture ...
Страница 134: ...5 14 Vaporizer manifold test fixture Anesthesia machine calibration device 1 MPa 10bar test pressure gauge ...
Страница 242: ...6 54 6 4 1 Electrical Connections Electrical Circuit Diagram ...
Страница 260: ...6 72 FOR YOUR NOTES ...
Страница 261: ...7 1 7 Parts 7 1 Main Unit 7 1 1 Main Unit 7 1 1 1 Exploded View ...
Страница 321: ......
Страница 322: ...P N 0621 20 78654 1 1 ...