3.9 Alarm Tests
3.9.1 Prepare for Alarm Tests
1. Connect a test lung or manual bag to the Y piece patient connection.
2. Set the bag/mechanical ventilation switch to the
position.
3. Set the system switch to the
position.
4. Set the system to Standby.
5. Set the ventilator controls as follows:
Ventilation mode: select the [
Vent Mode
] shortcut key and then [
VCV
].
[
TV
]: 500 ml.
[
Rate
]:12 BPM.
[
I:E
]:
1:2.
[
Plimit
]: 30 cmH
2
O.
[
PEEP
]: OFF.
6. Push the O
2
flush button to fill the bellows, folding bag rising to the top.
7. Turn the O
2
flow control to set the O
2
flow to 0.5 to 1 L/min.
8 Press
the key and select [
Ok
] from the pop-up menu to exit Standby.
9. Make sure that:
The ventilator displays the correct data.
The folding bag inside the bellows inflates and deflates normally during
mechanical ventilation.
Содержание WATO EX-55
Страница 1: ...WATO EX 55 65 Anesthesia Machine Service Manual ...
Страница 2: ......
Страница 4: ...II FOR YOUR NOTES ...
Страница 6: ...IV FOR YOUR NOTES ...
Страница 12: ...6 FOR YOUR NOTES ...
Страница 18: ...2 2 2 2 Gas Flow 2 2 1 Pneumatic Circuit Diagram vaporizer vaporizer Patient Waste gas disposal Patient Air ...
Страница 58: ...3 22 FOR YOUR NOTES ...
Страница 133: ...5 13 Negative pressure ball Circuit adapter test fixture Flow sensor pressure sampling pipeline test fixture ...
Страница 134: ...5 14 Vaporizer manifold test fixture Anesthesia machine calibration device 1 MPa 10bar test pressure gauge ...
Страница 242: ...6 54 6 4 1 Electrical Connections Electrical Circuit Diagram ...
Страница 260: ...6 72 FOR YOUR NOTES ...
Страница 261: ...7 1 7 Parts 7 1 Main Unit 7 1 1 Main Unit 7 1 1 1 Exploded View ...
Страница 321: ......
Страница 322: ...P N 0621 20 78654 1 1 ...