6-8
6.2.6 Remove the Throttling Device
1. Follow the steps described in
6.2.5Remove the Left Front Panel Assembly.
2. Disconnect the gas-in tube, gas-out tube and pressure sampling tube.
3. Open the cable ties and disconnect the connection lines from the flowmeter control
board. Unscrew the fixed screws.
Gas-in tube
Gas-out tube
Pressure sampling tube
Remove the
connection lines
Open the cable ties
Fixed screws