5-43
The seal for the bag arm is
damaged.
Replace the seal for the bag arm. It is
required to replace the seal once a
year.
The water collection cup gets
loose.
Check and tighten the water collection
cup.
The seal for the water collection
cup assembly is damaged.
Replace the seal for the water
collection cup assembly. It is required
to replace the seal once a year.
The patient circuit is separated
from the circuit adapter.
Reinstall the patient circuit. Ensure
that the circuit block is in the
(locked) position.
The seal for the circuit adapter
assembly is damaged.
Replace the seal, which is required to
be replaced once a year.
The bellows housing or folding
bag is not installed properly.
Re-install the bellows housing or
folding bag. Ensure their correct
installation.
The bellows sealing cushion
falls off or is damaged.
Replace the bellows sealing cushion,
which is required to be replaced once
a year.
The valve cover of the breathing
valve assembly is not installed
properly.
Re-install the valve cover and ensure
its correct installation.
The seal for the valve cover of
the breathing valve assembly is
damaged.
Replace the seal.
The O2 sensor is not installed
properly.
Re-install the O2 sensor and ensure its
correct installation.
The seal for the O2 sensor or the
seal for the O2 sensor plug is
damaged.
Replace the seal.
The seal underneath the
sodalime canister support is
damaged.
Replace the seal underneath the
sodalime canister support. It is
required to replace the seal once a
year.
The breathing tube connecting
the patient is damaged.
Replace the breathing tube.
The folding bag is damaged.
Replace the folding bag, which is
required to be replaced once a year.
Содержание WATO EX-55
Страница 1: ...WATO EX 55 65 Anesthesia Machine Service Manual ...
Страница 2: ......
Страница 4: ...II FOR YOUR NOTES ...
Страница 6: ...IV FOR YOUR NOTES ...
Страница 12: ...6 FOR YOUR NOTES ...
Страница 18: ...2 2 2 2 Gas Flow 2 2 1 Pneumatic Circuit Diagram vaporizer vaporizer Patient Waste gas disposal Patient Air ...
Страница 58: ...3 22 FOR YOUR NOTES ...
Страница 133: ...5 13 Negative pressure ball Circuit adapter test fixture Flow sensor pressure sampling pipeline test fixture ...
Страница 134: ...5 14 Vaporizer manifold test fixture Anesthesia machine calibration device 1 MPa 10bar test pressure gauge ...
Страница 242: ...6 54 6 4 1 Electrical Connections Electrical Circuit Diagram ...
Страница 260: ...6 72 FOR YOUR NOTES ...
Страница 261: ...7 1 7 Parts 7 1 Main Unit 7 1 1 Main Unit 7 1 1 1 Exploded View ...
Страница 321: ......
Страница 322: ...P N 0621 20 78654 1 1 ...