6-20
6.2.16 Remove the Three-way Valve
Disconnect the connection line between the three-way valve and the monitor board.
Disconnect the four pressure sampling tubes between the patient circuit and the three-way
valve and unscrew the two screws as shown below to remove the three-way valve.
6.2.17 Remove the Auxiliary O2 Supply Assembly
1. Unscrew the four screws on the rear panel of auxiliary O2 supply assembly, as shown
below.
Unscrew these two screws
Disconnect the pressure sampling
tubes between the patient circuit
and the three-way valve
Disconnect the pressure sampling tubes between the patient circuit and the three-way valve