5-41
9. Repeat steps 3 through 5 as shown below. If the vaporizer manifold assembly is leaky,
one or several rubber plain washers and seal of the vaporizer manifold assembly or its
contacted mechanical surface are damaged. Handle this problem as described in the
troubleshooting table.
II Ensure that the vaporizer manifold assembly with vaporizer mounted (mentioned in “I”)
is not leaky and perform the following operations.
1. Remove the vaporizer and mount the vaporizer manifold test fixture onto the
connector of the vaporizer manifold assembly (remove the seal between the connector
and the vaporizer when mounting the test fixture), as shown below.
2. Pull out No.25 PU tube which connects the float flowmeter to the vaporizer manifold
assembly. The end of the tube which connects the vaporizer manifold assembly is
pulled out but the other end is not.
3. Occlude the pulled-out tube end on the vaporizer manifold assembly by using
3126-06-00 tube plug.
Содержание WATO EX-55
Страница 1: ...WATO EX 55 65 Anesthesia Machine Service Manual ...
Страница 2: ......
Страница 4: ...II FOR YOUR NOTES ...
Страница 6: ...IV FOR YOUR NOTES ...
Страница 12: ...6 FOR YOUR NOTES ...
Страница 18: ...2 2 2 2 Gas Flow 2 2 1 Pneumatic Circuit Diagram vaporizer vaporizer Patient Waste gas disposal Patient Air ...
Страница 58: ...3 22 FOR YOUR NOTES ...
Страница 133: ...5 13 Negative pressure ball Circuit adapter test fixture Flow sensor pressure sampling pipeline test fixture ...
Страница 134: ...5 14 Vaporizer manifold test fixture Anesthesia machine calibration device 1 MPa 10bar test pressure gauge ...
Страница 242: ...6 54 6 4 1 Electrical Connections Electrical Circuit Diagram ...
Страница 260: ...6 72 FOR YOUR NOTES ...
Страница 261: ...7 1 7 Parts 7 1 Main Unit 7 1 1 Main Unit 7 1 1 1 Exploded View ...
Страница 321: ......
Страница 322: ...P N 0621 20 78654 1 1 ...