6-18
6.2.15 Disassemble the Expiratory Valve Assembly
1. Unscrew the screws fixing the connection line to the pneumatic block. Disconnect the
connection line between the monitor board and the pressure switch.
2. Disconnect the tubes.
Disconnect three pressure sampling tubes
Disconnect the tube from gas reservoir
Remove the drive gas tube
Remove the drive
gas tube
Disconnect the connection line
from the pressure switch
Disconnect the connection line
from the monitor board