6-25
6.2.22 Remove the Speaker
Disconnect the connection lines from the button board. Open the flat cable tie. Unscrew the
two screws to remove the speaker, as shown below.
6.2.23 Remove the Power Box
1. Remove the rear panel and disconnect the connection line from the auxiliary electrical
outlet.
Remove the
connection line
Flat cable tie
Disconnect the connection
line from the auxiliary
electrical outlet