4-15
4.2.6 Check the Pressure Sensor Accuracy
NOTE
z
Generally, measurement deviations do not easily occur to pressure sensors.
However, in case of maintaining or replacing the monitor board, three-way valve
assembly, or expiratory valve assembly, you need to perform pressure calibration
and check the flow sensors accuracy so as to confirm the effectiveness of
calibration.
To check the measurement accuracy of pressure sensors:
1. Make sure that the pressure sampling line and calibration device (or other pressure
measurement device) are connected in parallel, similar to tubes connection in pressure
calibration. For details, refer to
4.3.3Pressure Calibration (factory).
2. When the system is Standby, select the [Maintenance] shortcut key
→
[Factory
Maintenance >>]
→
[Diagnostic Test >>]
→
[Valves
—
Test Tool >>] to access the
[Valves
—
Test Tool] menu.
3. Set PEEP safety valve to [ON].
4. Increase the D/A value of the PEEP valve, causing the measured pressure value of the
anesthesia machine calibration device to fall with the following ranges respectively:
(5±1) cmH2O, (20±1) cmH2O, (50±1) cmH2O, (70±2) cmH2O, (90±2) cmH2O. Record
the measured pressure values of the airway pressure sensor and PEEP pressure sensor to
which each setting corresponds respectively.
5. Make sure that the deviation between the measured data of the airway pressure sensor,
PEEP pressure sensor and that of the anesthesia machine calibration device must not
exceed 1 cmH2O or 2% of the measured value of the calibration device, whichever is
greater. Otherwise, refer to
4.3.3Pressure Calibration (factory)
to perform pressure
calibration again.
Содержание WATO EX-55
Страница 1: ...WATO EX 55 65 Anesthesia Machine Service Manual ...
Страница 2: ......
Страница 4: ...II FOR YOUR NOTES ...
Страница 6: ...IV FOR YOUR NOTES ...
Страница 12: ...6 FOR YOUR NOTES ...
Страница 18: ...2 2 2 2 Gas Flow 2 2 1 Pneumatic Circuit Diagram vaporizer vaporizer Patient Waste gas disposal Patient Air ...
Страница 58: ...3 22 FOR YOUR NOTES ...
Страница 133: ...5 13 Negative pressure ball Circuit adapter test fixture Flow sensor pressure sampling pipeline test fixture ...
Страница 134: ...5 14 Vaporizer manifold test fixture Anesthesia machine calibration device 1 MPa 10bar test pressure gauge ...
Страница 242: ...6 54 6 4 1 Electrical Connections Electrical Circuit Diagram ...
Страница 260: ...6 72 FOR YOUR NOTES ...
Страница 261: ...7 1 7 Parts 7 1 Main Unit 7 1 1 Main Unit 7 1 1 1 Exploded View ...
Страница 321: ......
Страница 322: ...P N 0621 20 78654 1 1 ...