6-42
3. Pull off the sodalime canister from the lifting device.
WARNING
z
Sodalime is a caustic substance and is a strong irritant to eyes, skin and
respiratory system. Affected parts should be flushed with water. If irritation
continues after flushed by water, seek medical assistance immediately.
6.3.10 Remove the Water Collection Cup
1. Hold the water collection cup and turn it clockwise.
2. Remove the water collection cup.