3-19
b.Take out the nut, fixed plate and filter screen by turn.
3. The waste gas disposal system is not working or the pump rate is less than 60 L/min at
which the AGSS works normally. Check if the waste gas disposal system reaches the
pump rate range of 50-80 L/min specified by the AGSS.
3.10.2 Check the Transfer Tube and Active Scavenging Tube
Disconnect the tubes from other components.
1. Check the transfer tube and its connectors for damage. If any damage is detected,
replace the tube (four transfer tubes are available in the AGSS accessories for your
replacement).
2. Check:
1).The receiving hose and its connectors for damage.
2).If the connections between the receiving hose and its connectors are loose.
3).The nylon pad and seal for damage.
If any damage or loose connection is detected, replace the corresponding component.
Nut
Fixed plate
Filter
screen
Female 30 mm conical connector
Transfer tube
Male 30 mm conical connector.
Содержание WATO EX-55
Страница 1: ...WATO EX 55 65 Anesthesia Machine Service Manual ...
Страница 2: ......
Страница 4: ...II FOR YOUR NOTES ...
Страница 6: ...IV FOR YOUR NOTES ...
Страница 12: ...6 FOR YOUR NOTES ...
Страница 18: ...2 2 2 2 Gas Flow 2 2 1 Pneumatic Circuit Diagram vaporizer vaporizer Patient Waste gas disposal Patient Air ...
Страница 58: ...3 22 FOR YOUR NOTES ...
Страница 133: ...5 13 Negative pressure ball Circuit adapter test fixture Flow sensor pressure sampling pipeline test fixture ...
Страница 134: ...5 14 Vaporizer manifold test fixture Anesthesia machine calibration device 1 MPa 10bar test pressure gauge ...
Страница 242: ...6 54 6 4 1 Electrical Connections Electrical Circuit Diagram ...
Страница 260: ...6 72 FOR YOUR NOTES ...
Страница 261: ...7 1 7 Parts 7 1 Main Unit 7 1 1 Main Unit 7 1 1 1 Exploded View ...
Страница 321: ......
Страница 322: ...P N 0621 20 78654 1 1 ...