4-46
4.4.1.3 System Software Upgrade
NOTE
z
Before upgrading the system software, check the version information of the booting
software. If it is not the latest, upgrade the booting software to the latest version
first and make sure of software compatibility.
z
Before upgrading the system software, record the current settings of the anesthesia
machine so as to restore the pre-upgrade settings after software upgrade.
z
When selecting the upgrade package, make sure that the checksum and version are
same to that provided by the factory. You also need to check the machine code
corresponding to the anesthesia machine to be upgraded.
Before upgrading the system software, note the matching between machine name and
machine code as listed below.
Machine name
Machine code Notes
WATO EX-55
0621
/
WATO EX-65
0621
/
WATO EX-50 (2
nd
generation circuit
)
0613
Borrows the breathing circuit of WATO EX-55/65
WATO EX-60 (2
nd
generation circuit
)
0613
Borrows the breathing circuit of WATO EX-55/65
WATO EX-50 (improved
circuit)
0611C
Built-in dual-flow sensor circuit after EBU017
engineering change
WATO EX-60 (improved
circuit)
0611C
Built-in dual-flow sensor circuit after EBU017
engineering change
WATO EX-50 (1
st
generation circuit)
0611
External single-flow sensor circuit before EBU017
engineering change
WATO EX-60 (1
st
generation circuit)
0611
External single-flow sensor circuit before EBU017
engineering change
To upgrade the system software:
1. Check and confirm that the booting software of the anesthesia machine is of the latest
version. If not, refer to
4.4.1.2Booting Software Upgrade
to upgrade to the latest
version.
Содержание WATO EX-55
Страница 1: ...WATO EX 55 65 Anesthesia Machine Service Manual ...
Страница 2: ......
Страница 4: ...II FOR YOUR NOTES ...
Страница 6: ...IV FOR YOUR NOTES ...
Страница 12: ...6 FOR YOUR NOTES ...
Страница 18: ...2 2 2 2 Gas Flow 2 2 1 Pneumatic Circuit Diagram vaporizer vaporizer Patient Waste gas disposal Patient Air ...
Страница 58: ...3 22 FOR YOUR NOTES ...
Страница 133: ...5 13 Negative pressure ball Circuit adapter test fixture Flow sensor pressure sampling pipeline test fixture ...
Страница 134: ...5 14 Vaporizer manifold test fixture Anesthesia machine calibration device 1 MPa 10bar test pressure gauge ...
Страница 242: ...6 54 6 4 1 Electrical Connections Electrical Circuit Diagram ...
Страница 260: ...6 72 FOR YOUR NOTES ...
Страница 261: ...7 1 7 Parts 7 1 Main Unit 7 1 1 Main Unit 7 1 1 1 Exploded View ...
Страница 321: ......
Страница 322: ...P N 0621 20 78654 1 1 ...