6-40
6.3.7 Disassemble the Expiratory Check Valve Assembly
1. Turn the check valve cover counterclockwise to remove it.
2. Pull out the check valve.
The following picture shows the appearance of expiratory check valve assembly.
6.3.8 Disassemble the Inspiratory Check Valve Assembly
For details about how to disassemble the inspiratory check valve assembly, refer to
6.3.7Disassemble the Expiratory Check Valve Assembly
。