6-23
6.2.19 Remove the Main Control Board
Disconnect the connection lines from the main control board. Unscrew the 4 screws as
shown below to remove the main control board.
6.2.20 Remove the Button Board
The button board is located beside the battery box and main control board. Disconnect the
connection lines from the button board. Unscrew the four screws as shown below to remove
the button board.