4-25
9. Before calibration, make sure that the supply gas pressure is sufficient. If cylinder
supply is used, turn up the cylinder yoke (not cylinder regulator) enough before
calibration so as to ensure that the pressure reading on the O2 pressure gauge is kept
above 0.3 MPa. If pressure falls, turn up the cylinder yoke further.
10. Make sure that the anesthesia machine is in standby mode.
11. Turn off all fresh gases.
12. Select the [
Maintenance
] shortcut key
→
[
Factory Maintenance >>
]
→
enter the
required password
→
[
Factory Cal. >>
]
→
[
Flow Cal. >>
]
→
[
Start
].
13. After flow calibration success is prompted, refer to
4.2.5Check the Flow Sensor
Accuracy
to test the effectiveness of flow calibration. In case of calibration failure, first
fix the problem and then perform flow calibration again.
NOTE
z
After flow calibration, check the accuracy of flow sensors by referring to
4.2.5Check the Flow Sensor Accuracy
z
In case of calibration failure, first fix the problem and then perform flow
calibration again.
Cylinder yoke
Cylinder regulator
Содержание WATO EX-55
Страница 1: ...WATO EX 55 65 Anesthesia Machine Service Manual ...
Страница 2: ......
Страница 4: ...II FOR YOUR NOTES ...
Страница 6: ...IV FOR YOUR NOTES ...
Страница 12: ...6 FOR YOUR NOTES ...
Страница 18: ...2 2 2 2 Gas Flow 2 2 1 Pneumatic Circuit Diagram vaporizer vaporizer Patient Waste gas disposal Patient Air ...
Страница 58: ...3 22 FOR YOUR NOTES ...
Страница 133: ...5 13 Negative pressure ball Circuit adapter test fixture Flow sensor pressure sampling pipeline test fixture ...
Страница 134: ...5 14 Vaporizer manifold test fixture Anesthesia machine calibration device 1 MPa 10bar test pressure gauge ...
Страница 242: ...6 54 6 4 1 Electrical Connections Electrical Circuit Diagram ...
Страница 260: ...6 72 FOR YOUR NOTES ...
Страница 261: ...7 1 7 Parts 7 1 Main Unit 7 1 1 Main Unit 7 1 1 1 Exploded View ...
Страница 321: ......
Страница 322: ...P N 0621 20 78654 1 1 ...