4-19
4 Electronic
flowmeter
calibration
(factory)
Calibrate the electronic
flowmeter board.
1. The electronic flowmeter board is
replaced.
2. The throttling device of the electronic
flowmeter is replaced.
3. The deviation between the measured
value of the electronic flowmeter and that
of the standard flow measurement device
exceeds more than 10% of the reading or
0.5 L/min, whichever is greater.
5
Pressure and flow
zeroing (factory)
Calibrate the deviation
from zero point of the
monitor board and
auxiliary monitor board.
Flow or Paw waveforms deviates from the
baseline.
6 Electronic
flowmeter zeroing
(factory)
Calibrate the deviation
from zero point of the
electronic flowmeter
board.
The electronic flowmeter has a zero point
error. The electronic flowmeter still
displays flow when fresh gases are all
turned off.
7 O2
sensor
calibration (user)
Calibrate the accuracy of
O2 sensor at 21% and
100% O2.
1. The measured value of the O2 sensor
has a great deviation. The deviation
exceeds 3% both in Air and pure O2.
2. The O2 sensor is replaced.
3. The monitor board is replaced.
8 CO2
calibration
(factory)
Calibrate to cause the
module to work more
accurately.
The measurement deviation of the module
exceeds the specified accuracy range.
9 AG
calibration
(factory)
Calibrate to cause the
module to work more
accurately.
The measurement deviation of the module
exceeds the specified accuracy range.
10 O2
module
calibration
(factory)
Calibrate to cause the
module to work more
accurately.
1. The measurement deviation of the
module exceeds the specified accuracy
range.
2. The O2 module is replaced.
Select the [
Maintenance
] shortcut key
→
[
Factory Maintenance >>
]. Enter the required
password to access the [
Factory Maintenance
] menu, where you can perform the following
calibrations and settings.
Содержание WATO EX-55
Страница 1: ...WATO EX 55 65 Anesthesia Machine Service Manual ...
Страница 2: ......
Страница 4: ...II FOR YOUR NOTES ...
Страница 6: ...IV FOR YOUR NOTES ...
Страница 12: ...6 FOR YOUR NOTES ...
Страница 18: ...2 2 2 2 Gas Flow 2 2 1 Pneumatic Circuit Diagram vaporizer vaporizer Patient Waste gas disposal Patient Air ...
Страница 58: ...3 22 FOR YOUR NOTES ...
Страница 133: ...5 13 Negative pressure ball Circuit adapter test fixture Flow sensor pressure sampling pipeline test fixture ...
Страница 134: ...5 14 Vaporizer manifold test fixture Anesthesia machine calibration device 1 MPa 10bar test pressure gauge ...
Страница 242: ...6 54 6 4 1 Electrical Connections Electrical Circuit Diagram ...
Страница 260: ...6 72 FOR YOUR NOTES ...
Страница 261: ...7 1 7 Parts 7 1 Main Unit 7 1 1 Main Unit 7 1 1 1 Exploded View ...
Страница 321: ......
Страница 322: ...P N 0621 20 78654 1 1 ...