4-60
4. Set the bag/mechanical ventilation switch to the mechanical ventilation position.
5. Plug the Y piece into the test plug to close the breathing circuit.
6. Push the O2 flush button repeatedly to sweep the pointer across the pressure gauge.
7. Remove the Y piece from the test plug and release the O2 flush button. Check if the
pointer goes to zero.
8. Repeat the steps above if the pointer fails to go to zero. If the pointer still fails to go to
zero, replace the airway pressure gauge.
9. If the pointer goes to zero, re-install the lens onto the gauge. If the pointer still fails to go
to zero, replace the airway pressure gauge.
4.6 Adjust the APL Valve Accuracy
To calibrate the APL valve:
1. Connect the inspiration connector and bag arm port using a breathing tube, as shown
below.
Содержание WATO EX-55
Страница 1: ...WATO EX 55 65 Anesthesia Machine Service Manual ...
Страница 2: ......
Страница 4: ...II FOR YOUR NOTES ...
Страница 6: ...IV FOR YOUR NOTES ...
Страница 12: ...6 FOR YOUR NOTES ...
Страница 18: ...2 2 2 2 Gas Flow 2 2 1 Pneumatic Circuit Diagram vaporizer vaporizer Patient Waste gas disposal Patient Air ...
Страница 58: ...3 22 FOR YOUR NOTES ...
Страница 133: ...5 13 Negative pressure ball Circuit adapter test fixture Flow sensor pressure sampling pipeline test fixture ...
Страница 134: ...5 14 Vaporizer manifold test fixture Anesthesia machine calibration device 1 MPa 10bar test pressure gauge ...
Страница 242: ...6 54 6 4 1 Electrical Connections Electrical Circuit Diagram ...
Страница 260: ...6 72 FOR YOUR NOTES ...
Страница 261: ...7 1 7 Parts 7 1 Main Unit 7 1 1 Main Unit 7 1 1 1 Exploded View ...
Страница 321: ......
Страница 322: ...P N 0621 20 78654 1 1 ...