Acoustic Output Reporting Table (60601-2-37) C-17
Transducer Model: 35C50EA
Imaging Mode: B-mode /Smart3D/Tissue Harmonic Imaging
Index label
MI
TIS
TIB
TIC
At
surfac
e
Below
surfac
e
At
surfac
e
Below
surfac
e
Maximum index value
0.97
0.27
1.00
1.21
Index component value
0.27
0.27
1.00
0.26
Acoustic
Parameter
s
p
r,
α
at
z
MI
(MPa)
1.53
P
(mW)
126.00
126.00
152.00
P
1×1
(mW)
16.28
16.28
z
s
(cm)
/
z
b
(cm)
/
z
MI
(cm)
4.91
z
pii
,
α
(cm)
4.91
f
awf
(MHz)
2.46
3.42
3.39
2.46
Other
Informatio
n
prr
(Hz)
4,531.0
0
srr
(Hz)
55.00
n
pps
1.33
I
pa,
α
at
z
pii
,
α
(W/cm
2
)
202.84
I
spta,
α
at
z
pii
,
α
or
z
sii
,
α
(mW/cm
2
)
27.00
I
spta
at
z
pii
or
z
sii
(mW/cm
2
)
63.70
p
r
at
z
pii
(MPa)
2.23
Operating
control
conditions
Acoustic power
100%
100%
100%
100%
Display depth
14.8cm
14.8cm
14.8cm
14.8cm
Focus position
6.0cm
13.0cm
13.0cm
14.0cm
Working Frequency
H6.0
3.5
3.5
H6.0
PRF
4531
4531
4531
4531
Содержание DP-50 Exp Vet
Страница 2: ......
Страница 34: ...2 6 System Overview 2 6 Introduction of Each Unit Right View Left View...
Страница 42: ......
Страница 68: ......
Страница 128: ......
Страница 148: ......
Страница 166: ...10 18 DICOM For details on tast manager see 9 6 Animal Task Manager...
Страница 180: ......
Страница 220: ......
Страница 224: ......
Страница 236: ......
Страница 242: ......
Страница 248: ......
Страница 249: ...Acoustic Output Reporting Table 60601 2 37 C 1 Appendix C Acoustic Output Reporting Table 60601 2 37...
Страница 342: ...D 2 Printer Adapter Type Model SONY X898MD...
Страница 343: ...P N 046 017713 02 1 0...