Acoustic Output 14-3
TIC: Thermal index for applications, such as pediatric and adult cranial applications,
in which the ultrasound beam passes through bone near the beam entrance into the
body.
WFUMB (World Federation for Ultrasound in Medicine and Biology) guidelines: state
that temperature increase of 4 degree C for 5 min or more should be considered as
potentially hazardous to embryonic and fetal tissue.
The smaller the MI/TI values, the lower the bioeffects.
14.4.2 MI/TI Display
TI and MI values are displayed in the upper part of the screen in real-time. The operator
should monitor these index values during examinations and ensure that exposure time and
output values are maintained at the minimum amounts needed for effective diagnosis.
NOTE:
If there is a value of MI or TI exceeds 1.0, you must be careful to practice the
ALARA principle.
The display precision is 0.1
14.5 Acoustic Power Setting
Acoustic power adjustment
Click [A. power] in the menu to adjust the acoustic power percentage, and the value is
displayed on the screen. The greater the acoustic power percentage, the greater the current
acoustic output.
When the image is frozen, the system stops transmitting acoustic power.
Default setting of acoustic power
Selection of diagnostic applications is the most important factor for controlling ultrasound
output.
The permissible level of intensity of ultrasound differs depending on the region of interest.
For fetal examinations, in particular, much care must be exercised.
In this system, imaging setups can be created using the ultrasound output set by you. At this
time, the default function is disabled.
It is the user’s responsibility for any change to the
default settings.
Default choices
Initial power 7% to 100%*
*
Definition of 100%:
The maximum acoustic power of a transducer determined by the increase in
transducer surface temperature in the selected mode and the acoustic power
restrictions specified by the FDA.
NOTE:
This system automatically returns to the settings whenever changes are made to
the values (when you turn on the power, switch between probes, press [End Exam],
or select Return in the Setup menu). In the factory default settings, the Acoustic
Output is limited below settings. Following the ALARA restriction, you are allowed to
increase the acoustic power under FDA 510(k) Guidance-Track3 limits and to set it
in the image preset screen.
The acoustic output of the system has been measured and calculated in accordance with
IEC60601-2-37: 2015, FDA 510(K) GUIDANCE, IEC 62359: 2017, Ultrasonics-Field
Содержание DP-50 Exp Vet
Страница 2: ......
Страница 34: ...2 6 System Overview 2 6 Introduction of Each Unit Right View Left View...
Страница 42: ......
Страница 68: ......
Страница 128: ......
Страница 148: ......
Страница 166: ...10 18 DICOM For details on tast manager see 9 6 Animal Task Manager...
Страница 180: ......
Страница 220: ......
Страница 224: ......
Страница 236: ......
Страница 242: ......
Страница 248: ......
Страница 249: ...Acoustic Output Reporting Table 60601 2 37 C 1 Appendix C Acoustic Output Reporting Table 60601 2 37...
Страница 342: ...D 2 Printer Adapter Type Model SONY X898MD...
Страница 343: ...P N 046 017713 02 1 0...