C-12 Acoustic Output Reporting Table (60601-2-37)
Transducer Model: 35C20EA
Imaging Mode: PW-mode
Index label
MI
TIS
TIB
TIC
At
surfac
e
Below
surfac
e
At
surfac
e
Below
surfac
e
Maximum index value
0.42
0.37
0.47
1.31
Index component value
0.37
0.33
0.47
0.34
Acoustic
Parameter
s
p
r,
α
at
z
MI
(MPa)
0.67
P
(mW)
38.69
18.38
32.19
P
1×1
(mW)
31.20
14.82
z
s
(cm)
1.89
z
b
(cm)
1.66
z
MI
(cm)
1.00
z
pii
,
α
(cm)
1.00
f
awf
(MHz)
2.51
2.51
2.51
2.51
Other
Informatio
n
prr
(Hz)
699.00
srr
(Hz)
/
n
pps
/
I
pa,
α
at
z
pii
,
α
(W/cm
2
)
16.55
I
spta,
α
at
z
pii
,
α
or
z
sii
,
α
(mW/cm
2
)
16.97
I
spta
at
z
pii
or
z
sii
(mW/cm
2
)
20.18
p
r
at
z
pii
(MPa)
0.73
Operating
control
conditions
Acoustic power
100%
100%
100%
100%
Display depth
16.6cm
16.6cm
16.6cm
6.5cm
SV Position
14.0cm
16.0cm
10.0cm
4.0cm
Working Frequency
2.5
2.5
2.5
2.5
PRF
699
3900
6300
12001
SV
0.5mm
0.5mm
0.5mm
0.5mm
Содержание DP-50 Exp Vet
Страница 2: ......
Страница 34: ...2 6 System Overview 2 6 Introduction of Each Unit Right View Left View...
Страница 42: ......
Страница 68: ......
Страница 128: ......
Страница 148: ......
Страница 166: ...10 18 DICOM For details on tast manager see 9 6 Animal Task Manager...
Страница 180: ......
Страница 220: ......
Страница 224: ......
Страница 236: ......
Страница 242: ......
Страница 248: ......
Страница 249: ...Acoustic Output Reporting Table 60601 2 37 C 1 Appendix C Acoustic Output Reporting Table 60601 2 37...
Страница 342: ...D 2 Printer Adapter Type Model SONY X898MD...
Страница 343: ...P N 046 017713 02 1 0...