7-4 Measurement
7.4
Measurement Accuracy
Table 1 Error of 2D Images
Parameter
Value Range
Error
Distance
Full screen
Within ±3%; or when the measured value is less
than 40 mm, the error is less than1.5 mm.
Area (Trace)
Full screen
Within ±7%; or when the measured value is less
than 16 cm
2
, the error is less than 1.2 cm
2
.
Area (ellipse,
circle)
Full screen
Within ±7%; or when the measured value is less
than 16 cm
2
, the error is less than 1.2 cm
2
.
Angle
Full screen
Within ±3%.
Circumference
Full screen
Within ±10%.
Distance (iScape) Full screen
Micro-convex probe: within ±15%
Other probe: within ±10%
Table 2 Volume Measurements
Parameter
Value
Range
Error
Volume
Full screen
Within ±10 %; or when the measured value is less than 64 cm
3
,
the error is less than 6.4 cm
3
.
Table 3 Time/Motion Measurements
Parameter
Value Range
Error
Distance
Full screen
Within ±3 %; or when the measured value is less
than 40mm, the error is less than 1.5mm.
Time
Timeline Display
Within ±2%.
Heart rate
Timeline Display
Within ±4%.
PW velocity
10 cm/s ~200cm/s
When the correction angle is
≤60º, the PW
velocity ≤±10%.
NOTE:
Within the selected field range, the measurement accuracy is ensured within the
range mentioned above. The accuracy specifications are performance in the worst
conditions, or based on the real test for the system, regardless of acoustic speed
error.
Содержание DP-50 Exp Vet
Страница 2: ......
Страница 34: ...2 6 System Overview 2 6 Introduction of Each Unit Right View Left View...
Страница 42: ......
Страница 68: ......
Страница 128: ......
Страница 148: ......
Страница 166: ...10 18 DICOM For details on tast manager see 9 6 Animal Task Manager...
Страница 180: ......
Страница 220: ......
Страница 224: ......
Страница 236: ......
Страница 242: ......
Страница 248: ......
Страница 249: ...Acoustic Output Reporting Table 60601 2 37 C 1 Appendix C Acoustic Output Reporting Table 60601 2 37...
Страница 342: ...D 2 Printer Adapter Type Model SONY X898MD...
Страница 343: ...P N 046 017713 02 1 0...