5-10 Image Optimization
During M mode imaging, you can switch between B and M menu from the menu title.
During M Mode scanning, frequency and acoustic power of the transducer are
synchronous with that of B Mode.
Adjustment of the depth or TGC to the B Mode image will lead to corresponding changes
in M Mode image.
5.5.3
M Mode Image Optimization
Gain
Description
To adjust the gain of M mode image. The real-time gain value is displayed in the
image parameter area in the upper left corner of the screen.
Operation
Rotate the <Gain/ iTouch> knob clockwise to increase the gain, and anti-
clockwise to decrease.
The adjusting range is 0-100.
Effects
Increasing the gain will brighten the image and you can see more received
signals. However, noise may also be increased.
Depth
Description
This function is used to adjust the display depth of sampling, the real-time
value of which is displayed on the image parameter area in the upper left
corner of the screen.
Operation
Use the <Depth/Zoom> knob to adjust the depth.
The adjustable depth values vary depending upon the probe types.
Effects
Increase the depth to see tissue in deeper locations, while decrease the depth
to see tissue in shallower locations.
Impacts
Depth increase will cause a decrease in the frame rate.
Frequency
Description
This function is used to select the operating frequency of the current probe,
the real-time value of which is displayed in the image parameter area in the
upper left corner of the screen, where “F” represents B mode frequency, and
“FH” represents harmonic frequency.
Operation
Adjust it through [Frequency] on the image menu, wherein
“H” means the
harmonic frequency.
Values of frequency vary depending upon the probe types. Select the
frequency according to the detection depth and current tissue characteristics.
Effects
The higher the frequency the better the near field resolution but the worse the
force of penetration.
Harmonic imaging enhances near field resolution and reduces low-frequency
and large amplitude noise, so as to improve small parts imaging.
A. power
Description
Refers to the power of ultrasonic wave transmitted by the probe, the real-time
value of which is displayed in the image parameter area in the upper left corner
of the screen.
Содержание DP-50 Exp Vet
Страница 2: ......
Страница 34: ...2 6 System Overview 2 6 Introduction of Each Unit Right View Left View...
Страница 42: ......
Страница 68: ......
Страница 128: ......
Страница 148: ......
Страница 166: ...10 18 DICOM For details on tast manager see 9 6 Animal Task Manager...
Страница 180: ......
Страница 220: ......
Страница 224: ......
Страница 236: ......
Страница 242: ......
Страница 248: ......
Страница 249: ...Acoustic Output Reporting Table 60601 2 37 C 1 Appendix C Acoustic Output Reporting Table 60601 2 37...
Страница 342: ...D 2 Printer Adapter Type Model SONY X898MD...
Страница 343: ...P N 046 017713 02 1 0...