System Preparation 3-3
Checking after it is turned on
Press the power button in the upper right corner on the control panel.
To check the system after the system is turned on:
No.
Check Item
<1>
There shall be no unusual sounds or smells indicating possible
overheating.
<2>
There shall be no persistently displayed system error message.
<3>
There shall be no evident excessive noise, discontinuous, absent or black
artifacts in the B Mode image.
<4>
Check if there is abnormal heat on the surface of the probe during an
ultrasound procedure.
<5>
The control panel keys and knobs are fully functional.
<6>
The exam date and time are the same as the system date and time, and
are displayed correctly.
WARNING:
1.
If you use a probe giving off excessive heat, it may burn
the animal.
2.
If you find anything not functioning properly, this may
indicate that the system is defective. In this case, shut
down the system immediately and contact Mindray
Customer Service Department or sales representative.
NOTE:
When you start the system or switch between probes, you will hear clicking
sounds
– this is expected behavior.
3.3.2
Powering OFF
You need to follow the correct procedures to power off the system. In addition, after you
upgrade the software or when the system is down, you need to power off and restart it.
If you will not use the system for a long period of time, after powering off the system, you shall
disconnect the external power supply; turn off powers of all peripherals connected to the
system.
To power off your system normally:
Gently press the power button once on the upper right corner of the control panel. The
[Shutdown Confirm] screen appears. See the figure below.
Shut down: To power off the system normally.
Содержание DP-50 Exp Vet
Страница 2: ......
Страница 34: ...2 6 System Overview 2 6 Introduction of Each Unit Right View Left View...
Страница 42: ......
Страница 68: ......
Страница 128: ......
Страница 148: ......
Страница 166: ...10 18 DICOM For details on tast manager see 9 6 Animal Task Manager...
Страница 180: ......
Страница 220: ......
Страница 224: ......
Страница 236: ......
Страница 242: ......
Страница 248: ......
Страница 249: ...Acoustic Output Reporting Table 60601 2 37 C 1 Appendix C Acoustic Output Reporting Table 60601 2 37...
Страница 342: ...D 2 Printer Adapter Type Model SONY X898MD...
Страница 343: ...P N 046 017713 02 1 0...