Probes and Biopsy 12-35
2.
Separate the bracket and the transducer from the needle.
3.
Screw the pinch nut to release the needle-guided bracket.
4.
Separate the bracket and the transducer.
NGB-003
Metal/needle detachable needle-guided bracket:
(1) Screw the nut of the guiding block and remove the guiding block slightly along the
direction of the needle’s tail.
(2) Separate the residual part of the needle-guide bracket and the transducer from the
needle.
(3) Screw the pinch nut of the bracket, and remove the needle-guided bracket from the
transducer.
NGB-004
Hold the transducer in the left hand, unscrew the locking nut with the right hand to open
the retaining clamp, and then raise the needle-guided bracket to separate the locating
bulge from the locating grooves.