C-2 Acoustic Output Reporting Table (60601-2-37)
C.1
10L24EA
Transducer Model: 10L24EA
Imaging Mode: M-mode
Index label
MI
TIS
TIB
TIC
At
surfac
e
Below
surfac
e
At
surfac
e
Below
surfac
e
Maximum index value
1.07
0.20
0.40
0.31
Index component value
0.20
0.13
0.19
0.40
Acoustic
Parameter
s
p
r,
α
at
z
MI
(MPa)
2.89
P
(mW)
6.89
3.27
7.33
P
1×1
(mW)
6.89
3.27
z
s
(cm)
1.00
z
b
(cm)
1.08
z
MI
(cm)
1.08
z
pii
,
α
(cm)
1.08
f
awf
(MHz)
7.29
6.03
5.87
6.07
Other
Informatio
n
prr
(Hz)
1,000.0
0
srr
(Hz)
/
n
pps
/
I
pa,
α
at
z
pii
,
α
(W/cm
2
)
265.84
I
spta,
α
at
z
pii
,
α
or
z
sii
,
α
(mW/cm
2
)
44.69
I
spta
at
z
pii
or
z
sii
(mW/cm
2
)
80.30
p
r
at
z
pii
(MPa)
3.65
Operating
control
conditions
Acoustic power
100%
100%
100%
100%
Display depth
12.9cm
12.9cm
12.9cm
12.9cm
Focus position
2.0cm
5.0cm
2.0cm
5.5cm
Working Frequency
10
H12.0
H12.0
H12.0
PRF
1000
2999
2999
2999
Содержание DP-50 Exp Vet
Страница 2: ......
Страница 34: ...2 6 System Overview 2 6 Introduction of Each Unit Right View Left View...
Страница 42: ......
Страница 68: ......
Страница 128: ......
Страница 148: ......
Страница 166: ...10 18 DICOM For details on tast manager see 9 6 Animal Task Manager...
Страница 180: ......
Страница 220: ......
Страница 224: ......
Страница 236: ......
Страница 242: ......
Страница 248: ......
Страница 249: ...Acoustic Output Reporting Table 60601 2 37 C 1 Appendix C Acoustic Output Reporting Table 60601 2 37...
Страница 342: ...D 2 Printer Adapter Type Model SONY X898MD...
Страница 343: ...P N 046 017713 02 1 0...