12-28 Probes and Biopsy
4) Select a proper guiding block and push it into the groove above the angle block,
and clamp it tightly.
5) Screw the nut of the block to secure the block.
6) Insert a biopsy needle with the same specification as that of the guiding block
into the hole of the guiding block.
Plastic
1) Put on the sterile transducer sheath.
2) Hold the transducer by one hand, select proper needle-guided bracket, and
hold it with the other hand. Align the narrow end tab of the needle-guided
bracket with the groove of the transducer, then push the needle-guided bracket
forward, making the tabs and the grooves of the needle-guided bracket to
match with the grooves and tabs of the transducer.
3) Check manually to confirm that the needle-guided bracket is securely installed
on the transducer.
4) Select a proper guiding block and push it into the groove above the angle block,
and clamp it tightly.