DETAILED OPERATION - Section 4
MTS-1700 SERIES OPERATION AND REFERENCE MANUAL 4-3
CU A002 15A
MANTA TEST SYSTEMS
In normal application, the MTS-1710 begins in PREFAULT state. A start trigger action will change the
MTS-1710 to the FAULT state.
A start trigger action may include pressing the [FAULT] button, a transition on the [EXTERNAL START]
trigger inputs, an RS-232C STR command, or an auxiliary start signal from the optional digital input ports.
While in the FAULT state, all start trigger actions will be ignored.
Now a stop trigger action will change the MTS-1710 to the POSTFAULT state. A stop trigger action can
originate from a transition on the [EXTERNAL STOP] trigger inputs, and RS-232C STP command, or an
auxiliary stop signal from the optional digital input ports.
While in the POSTFAULT state, all further start and stop trigger actions are ignored, preventing the MTS-
1710 from re-entering a FAULT state.
Now, only a reset action will restore the MTS-1710 to the PREFAULT state. The reset action may be caused
by pressing the [STOP/RESET] button, or by the RS-232C RES command.
Changing from dynamic to static operation mode also will cause a reset action, returning the MTS-1710 to
the PREFAULT state if it was previously in the FAULT or POSTFAULT state.
4.2.3
Blinking Indicators in Dynamic Mode
Note that, in dynamic operation mode, one or more of the [PREFAULT], [FAULT] or [POSTFAULT]
buttons may be illuminated at a time. This is due to the fact that the prefault and postfault voltage/current
may be turned on or off (see Section 4.4).
If the prefault or postfault levels are armed to be on, its corresponding pushbutton will be illuminated. As a
visual aid, if the output levels are turned on in a particular state, and the MTS-1710 is in the corresponding
state, the corresponding button will blink.
For example, if the operation mode is dynamic, and the [PREFAULT] button is lit, the [POSTFAULT]
button isn’t lit, and the [FAULT] button is blinking, then the MTS-1710 is in the fault state, with live
outputs, prefault outputs are set to on, and postfault outputs are set to off.
4.3
CHARACTERISTICS OF FAULT STATES
PREFAULT
FAULT
POSTFAULT
V & I outputs at prefault settings
(or off if prefault set off)
V & I outputs at fault settings
V & I outputs at postfault
settings
(or off if postfault set off)
Timer=0.0sec
Timer running
Timer reading frozen
Normal reading speed
Fast reading speed
(Dynamic operation mode only) *
All readings frozen at their
value when stop trigger
occurred
AUX contacts open**
AUX contacts closed**
AUX contacts open**
Содержание MTS-1700 SERIES
Страница 30: ...TABLE OF CONTENTS xxx MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 38: ...INTRODUCTION Section 1 1 8 MTS 1700SERIESOPERATIONANDREFERENCEMANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 44: ...SPECIFICATIONS Section 2 2 6 MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 248: ...FAULT PLAYBACK Section 7 7 12 MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 262: ...SERVICING Section 8 8 14 MTS 1700 SERIES OPERATION ANDREFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 320: ...PROGRAMMABLE DIGITAL I O CHANNELS A 14 MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 340: ...E 4 MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 344: ...GLOSSARY F 4 MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...