![Manta MTS-1700 SERIES Скачать руководство пользователя страница 229](http://html1.mh-extra.com/html/manta/mts-1700-series/mts-1700-series_operation-and-reference-manual_1723616229.webp)
RS-232C INTERFACE - Section 6
MTS-1700 SERIES OPERATION AND REFERENCE MANUAL 6-43
CU A002 15A
MANTA TEST SYSTEMS
AVA#,#
Amplitude Va
•
Set individual Va amplitude
•
First argument is amplitude. Valid values are 0 - 150.00 volts
•
Second argument specifies the fault state. Valid values are 0 - 2.
0 - prefault Va
1 - fault Va
2 - postfault Va
•
This command is only valid in individual amplitude adjustment mode.
AVB#,#
Amplitude Vb
•
Set individual Vb amplitude
•
First argument is amplitude. Valid values are 0 - 150.00 volts
•
Second argument specifies the fault state. Valid values are 0 - 2.
0 - prefault Vb
1 - fault Vb
2 - postfault Vb
•
This command is only valid in individual amplitude adjustment mode.
AVC#,#
Amplitude Vc
•
Set individual Vc amplitude
•
First argument is amplitude. Valid values are 0 - 150.00 volts.
•
Second argument specifies the fault state. Valid values are 0 - 2.
0 - prefault Vc
1 - fault Vc
2 - postfault Vc
•
This command is only valid in individual amplitude adjustment mode.
AI1#,#
Amplitude I1 or I3
•
Set individual I1 or I3 amplitude
•
First argument is amplitude. Valid values are:
0 - 30.000 amps for I1-LOW or I3 current modes
0 - 60.000 amps for I1&I2 current mode
0 - 90.000 amps for I1-HIGH current mode
•
Second argument specifies the fault state. Valid values are 0 - 2.
0 - prefault I1 (or I3)
1 - fault I1 (or I3)
2 - postfault I1 (or I3)
•
This command is only valid in individual amplitude adjustment mode.
AI2#,#
Amplitude I2
•
Set individual I2 amplitude
•
First argument is amplitude. Valid values are 0 - 30.0 amps in I1&I2 mode or 0 -
17.0 in I2-HARMONIC.
•
Second argument specifies the fault state. Valid values are 0 - 2.
0 - prefault I2
1 - fault I2
2 - postfault I2
•
This command is only valid in individual amplitude adjustment mode.
•
This command is only valid in I1&I2 and I2-HARMONIC current modes.
Содержание MTS-1700 SERIES
Страница 30: ...TABLE OF CONTENTS xxx MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 38: ...INTRODUCTION Section 1 1 8 MTS 1700SERIESOPERATIONANDREFERENCEMANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 44: ...SPECIFICATIONS Section 2 2 6 MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 248: ...FAULT PLAYBACK Section 7 7 12 MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 262: ...SERVICING Section 8 8 14 MTS 1700 SERIES OPERATION ANDREFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 320: ...PROGRAMMABLE DIGITAL I O CHANNELS A 14 MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 340: ...E 4 MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 344: ...GLOSSARY F 4 MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...