SERVICING - Section 8
MTS-1700 SERIES OPERATION AND REFERENCE MANUAL 8-11
CU A002 15A
MANTA TEST SYSTEMS
Select OTHER SERV CALIBRATION MEASUREMENT I3 in the menu. Turn the [MODIFY] knob
to adjust the calibration factor until the reading is the same as read on a reference meter (the output
current doesn’t change when this is done).
Press [SELECT].
7.
Turn the current down to zero. Remove the short from the external start trigger inputs.
8.
Set the dynamic measurement mode to PEAK (see section 5.1.1), and the operation mode to
DYNAM IC. Press [FAULT].
Repeat steps 3 to 7 to adjust the peak measurement calibration factors.
IMPORTANT: This must be done with the dynamic measurement mode set to PEAK and in the
FAULT state (FAULT button flashing).
9.
Repeat steps 3-8 for IBC this time by selecting
Φ−Φ
B-C fault mode.
10. Repeat steps 3-8 for ICA this time by selecting
Φ−Φ
C-A fault mode.
11. When done, the CAL1 command must be sent via the RS-232C interface to enable saving of the
calibration factors.
Next, select OTHER SERV CALIBRATION SAVE to save the calibration factors in non-volatile
memory so that these new calibration factors will be remembered next time the MTS-1720 is powered
up.
8.6 DC VOLTAGE MEASUREMENT CALIBRATION
This applies only to units with the DC voltage option installed.
1. Press [VOLTAGE] once or twice, as required, to display DC voltage on the [VALUE DISPLAY].
Connect an external reference DC voltmeter to the [DC VOLTS OUTPUT] terminals.
2. Select SETTINGS DC-VOLTS CONTINUOUS-ADJ in the menu. Turn the [MODIFY] knob to adjust
the voltage output to maximum (300V). Press SELECT.
3. Press the MODE/MENU button to exit the menu. Press the DC Voltage enable button.
4. Select OTHER SERV CALIBRATION MEASUREMENT DC-VOLTS in the menu. Turn the
[MODIFY] knob to adjust the calibration factor until the reading is the same as read on a reference meter
(the output voltage doesn’t change when this is done).
Press [SELECT].
5. When done, the CAL1 command must be sent via the RS-232C interface to enable saving of the
calibration factors.
Содержание MTS-1700 SERIES
Страница 30: ...TABLE OF CONTENTS xxx MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 38: ...INTRODUCTION Section 1 1 8 MTS 1700SERIESOPERATIONANDREFERENCEMANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 44: ...SPECIFICATIONS Section 2 2 6 MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 248: ...FAULT PLAYBACK Section 7 7 12 MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 262: ...SERVICING Section 8 8 14 MTS 1700 SERIES OPERATION ANDREFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 320: ...PROGRAMMABLE DIGITAL I O CHANNELS A 14 MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 340: ...E 4 MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 344: ...GLOSSARY F 4 MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...