MTS-1750 HIGH CURRENT SOURCE - Section 10
MTS-1700 SERIES OPERATION AND REFERENCE MANUAL 10-11
CU A002 15A
MANTA TEST SYSTEMS
Otherwise, the operation is as described in Sections 5.12.2.1.1 and 6.7.1.
For the MTS-1710 + MTS-1720 + (3) MTS-1750 configuration with a single phase load, the current
amplitude is programmed by the AI1#,# command (450A maximum), and the current phase angle is always
fixed at 0°. Otherwise, the operation is as described in Sections 5.12.2.1.4 and 6.7.1.
For the MTS-1710 + MTS-1720 + (3) MTS-1750 configuration with a three phase load, the current
amplitudes are programmed by the AIA#,# AIB#,# and AIC#,# commands (150A maximum), and the
current phase angles are programmed by the PHS#,#,4 PHS#,#,5 and PHS#,#,6 commands. Otherwise, the
operation is as described in Sections 5.12.2.1.5 and 6.7.2.
10.4.4.2 FAULT PLAYBACK WITH THE MTS-1750.
Fault playback can be performed with the MTS-1750 in any of the described configurations.
For the MTS-1710 + MTS-1750 configuration, the maximum current is programmed by the AIA command
(150A maximum). Otherwise, the operation is as described in Sections 7.3.1.1 and 7.5.1.
For the MTS-1710 + MTS-1720 + (3) MTS-1750 configuration with a single phase load, the maximum
current is programmed by the AIA command (450A maximum). Otherwise, the operation is as described in
Sections 7.3.1.1 and 7.5.1.
For the MTS-1710 + MTS-1720 + (3) MTS-1750 configuration with a three phase load, the maximum
current for the A, B and C channels is programmed by the AIA, AIB and AIC commands respectively (150A
maximum). Otherwise, the operation is as described in Sections 7.3.1.2 and 7.5.2.
10.4.4.3 PARALLELLING MORE THAN TWO MTS-1710/MTS-1720 SYSTEMS WITH
MTS-1750s.
Multiple MTS-1710s with or without MTS-1720s can be paralleled together to provide single phase or
three-phase high current, as described in Section 4.6.10.6. Any of these configurations further can be
enhanced by adding one or more MTS-1750s to provide even higher current output.
10.4.4.4 OPERATION OF THE MTS-1750 AS A STAND-ALONE CURRENT SOURCE.
The MTS-1750 can be used as a stand alone high current source without the need of a 1710. The
configuration requires a function generator used to control the amplitude, frequency and shape of the
generated signal.
Connection and Power
•
Turn on the Function Generator and ensure the generator has a minimum output voltage.
•
Select the type of signal (sine wave, square wave, etc.) which you wish to amplify.
•
Turn off the Function Generator, and ensure the MTS-1750 is also turned off.
•
Connect the leads of the Function Generator to pins 2 and 9, as shown in Figure 10.9.
•
Short together pin 12 (ground) to pin 8 (Amplifier enable). Pin 8 is the amplifier enable and is active
low. Shorting this pin to ground forces will always enable the amplifiers.
Содержание MTS-1700 SERIES
Страница 30: ...TABLE OF CONTENTS xxx MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 38: ...INTRODUCTION Section 1 1 8 MTS 1700SERIESOPERATIONANDREFERENCEMANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 44: ...SPECIFICATIONS Section 2 2 6 MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 248: ...FAULT PLAYBACK Section 7 7 12 MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 262: ...SERVICING Section 8 8 14 MTS 1700 SERIES OPERATION ANDREFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 320: ...PROGRAMMABLE DIGITAL I O CHANNELS A 14 MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 340: ...E 4 MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 344: ...GLOSSARY F 4 MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...