SERVICING - Section 8
MTS-1700 SERIES OPERATION AND REFERENCE MANUAL 8-5
CU A002 15A
MANTA TEST SYSTEMS
1. Set the DC voltage to 24VDC by selecting MENU SETTINGS DCVOLTS 24V.
2. Connect a DC Volt reference meter to the DC volt outputs.
3. Exit MENU and press the DC VOLTAGE enable button.
4. Select OTHER SERV CALIBRATION GENERATION VDC_LOW in the menu. Turn the [MODIFY]
knob to adjust the calibration factor until the reference meter reads 24VDC ±0.24V.
Press [SELECT]
5.
Press [PREVIOUS] several times to reach the main menu.
6. Select SETTINGS DCVOLTS 250. Exit the MENU.
7.
Select OTHER SERV CALIBRATION GENERATION VDC_HIGH in the menu.Turn the [MODIFY]
knob to adjust the calibration factor until the reference meter reads 250VDC ±2.5V.
Press [SELECT]
8.
Select OTHER SERV CALIBRATION SAVE in the CAL menu to save the calibration factors in non-
volatile memory so that these new calibration factors will be remembered next time the MTS-1720 is
powered up. Or, send the CAL3 command via RS232 interface.
8.5 MEASUREMENT CALIBRATION
The recommended calibration interval is one year. Front panel calibration of V & I measurements can be
done using the menu.
There are three voltage scales, two current scales for I1, and two current scales for I1-HIGH, with a separate
calibration factor for each.
AC Voltage scales (approx)
0-82V, 82V-328V
I1 scales (approx)
0-8.2A, 8.2A-32.8A
I1-HIGH scales (approx)
0-24.5A, 24.5A-98A
I3 scale
Phase to neutral: 0-8.2A, 8.2-32.8A, 25-45A
Phase to Phase: 0-16.4A, 16.4-65.6A
I4 scale (approx)
0-10A
DC Voltage scale
0-300V
For the AC voltage, I1, I1-HIGH and I3 measurements, there are both RMS and PEAK measurement mode
calibration factors. Calibration of these scales must be done once with dynamic measurement mode set to
RMS, and then again in FAULT state, with dynamic measurement mode set to PEAK (see Section 5.7).
Before starting the calibration procedures, the CAL1 command must be sent via the RS-232 interface to
enable saving of the calibration factors.
A 0.1% accurate external reference meter is required to perform the calibration.
Содержание MTS-1700 SERIES
Страница 30: ...TABLE OF CONTENTS xxx MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 38: ...INTRODUCTION Section 1 1 8 MTS 1700SERIESOPERATIONANDREFERENCEMANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 44: ...SPECIFICATIONS Section 2 2 6 MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 248: ...FAULT PLAYBACK Section 7 7 12 MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 262: ...SERVICING Section 8 8 14 MTS 1700 SERIES OPERATION ANDREFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 320: ...PROGRAMMABLE DIGITAL I O CHANNELS A 14 MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 340: ...E 4 MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 344: ...GLOSSARY F 4 MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...