![Manta MTS-1700 SERIES Скачать руководство пользователя страница 212](http://html1.mh-extra.com/html/manta/mts-1700-series/mts-1700-series_operation-and-reference-manual_1723616212.webp)
RS-232C INTERFACE - Section 6
6-26 MTS-1700 SERIES OPERATION AND REFERENCE MANUAL
MANTA TEST SYSTEMS
CU A002 15A
!3
<- data point at 2 degrees
!4
!5
<- data point at 3 degrees
!5
..etc.
(note that some points are repeated due to 8-bit amplitude resolution)
.
.
!252
<- data point at 178 degrees
!252
<- data point at 178.5 degrees
!253
<- data point at 179 degrees
!254
!255
<- data point at 180 degrees (+Max output)
!255
!254
<- data point at 181 degrees
!253
!253
<- data point at 182 degrees
..etc.
(triangular waveform definition continued)
.
.
!5
<- data point at 356.5 degrees
!5
<- data point at 357 degrees
!4
!3
<- data point at 358 degrees
!3
!2
<- data point at 359 degrees
!1
<- last data point at 359.5 degrees
To enable the programmable waveform, send the following commands:
PWC1
<- enable programmable waveform
RCR18.6
<- rate=18.6kHz, filter at 9.3kHz
LOC
<- local control mode
Now the amplitude, phase and frequency of the waveform may be controlled from the front panel (or RS-
232C interface, if desired).
To return to normal sine wave output, send the following commands:
REM
<- remote control mode
PWC0
<- programmable waveform off
LOC
<- return to local control mode
6.2.14.4 EXAMPLE - 14 BIT DATA.
The following is an example usage of the programmable waveform commands which redefine the sine
wave output to a triangular wave.
Содержание MTS-1700 SERIES
Страница 30: ...TABLE OF CONTENTS xxx MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 38: ...INTRODUCTION Section 1 1 8 MTS 1700SERIESOPERATIONANDREFERENCEMANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 44: ...SPECIFICATIONS Section 2 2 6 MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 248: ...FAULT PLAYBACK Section 7 7 12 MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 262: ...SERVICING Section 8 8 14 MTS 1700 SERIES OPERATION ANDREFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 320: ...PROGRAMMABLE DIGITAL I O CHANNELS A 14 MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 340: ...E 4 MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 344: ...GLOSSARY F 4 MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...