OPERATION SUMMARY - Section 3
3-16 MTS-1700 SERIES OPERATION AND REFERENCE MANUAL
MANTA TEST SYSTEMS
CU A002 15A
2. Press [STATIC/DYNAMIC], as required, to select STATIC operation mode.
3. Press and hold [FAULT], or short the [EXTERNAL START] contact inputs. This holds the MTS-1710
in the fault state, indicated by the illuminated [FAULT] button.
4. Press [VOLTAGE] to display/adjust voltage.
5. Turn the [MODIFY] knob to adjust the voltage until the relay operates. When the relay contacts close,
the [TONE] button will light. If the audible stop trigger tone has been enabled (by pressing the
[TONE] button), the tone also will sound. The voltage may be adjusted, as required, to determine the
pickup level.
6. Release the [FAULT] button, and remove connections to the [EXTERNAL START] inputs.
3.3.6.2 TIMING TEST.
To check the operation time:
1. Perform steps 1-3 of the pickup check above.
2. Press [VOLTAGE] to display/adjust voltage. Set the desired overvoltage or undervoltage level.
3. Release the [FAULT] button, and remove connections to the [EXTERNAL START] inputs.
4. If prefault voltage is desired, turn [PREFAULT] on, and set desired prefault voltage.
5. Press [STATIC/DYNAMIC] to select DYNAMIC operation mode. Press [FAULT] to initiate.
6. When the relay operates, the [STOP/RESET] button will illuminate. Press [TIME] to read the operate
time. Press [TIME] again to toggle between time in seconds/cycles. Press [STOP/RESET] to return to
the prefault state, and to prepare for the next test.
7. For each new value of voltage, repeat steps 2 to 6.
3.3.6.3 TARGET/SEAL-IN TEST.
To check the target and seal-in features of a DC current-operated target:
1. Make test connections, as shown in Figure 3.7. Select I4-DC current mode, and static operation mode.
Select
Φ
-N/A-N fault mode. If more than 150V is required, connect the relay across A-B and select
2
Φ
-N/A-B fault mode.
2. Turn [PREFAULT] on. Press [VOLTAGE] and turn the [MODIFY] knob to set the prefault voltage, as
required.
3. Press and hold the [FAULT] button, and turn the [MODIFY] knob to set the fault voltage. Release the
[FAULT] button.
4. If you want a value other than the default voltage during postfault, press [POSTFAULT], and set the
postfault voltage.Then turn [PREFAULT] on. To verify the seal-in function, set the postfault voltage to
a value which doesn’t operate the relay.
5. Use the current preset feature to preset the DC current (usually 0.2A or 2.0A). Note that the [MODE/
MENU DISPLAY] should indicate DC-AMPS while making this setting.
6. Select dynamic operation mode. Turn [POSTFAULT] on. (This is for DC current to be on in postfault
to operate the target and seal-in).
7. Press [FAULT] to initiate.When the relay operates, the MTS-1710 will change to postfault state, the
target and seal-in should operate, and the fault voltage level will be removed.
8. Turn [POSTFAULT] off to remove the DC current, and to release the seal-in unit.
Содержание MTS-1700 SERIES
Страница 30: ...TABLE OF CONTENTS xxx MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 38: ...INTRODUCTION Section 1 1 8 MTS 1700SERIESOPERATIONANDREFERENCEMANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 44: ...SPECIFICATIONS Section 2 2 6 MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 248: ...FAULT PLAYBACK Section 7 7 12 MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 262: ...SERVICING Section 8 8 14 MTS 1700 SERIES OPERATION ANDREFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 320: ...PROGRAMMABLE DIGITAL I O CHANNELS A 14 MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 340: ...E 4 MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 344: ...GLOSSARY F 4 MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...