DETAILED OPERATION - Section 4
4-46
MTS-1700 SERIES OPERATION AND REFERENCE MANUAL
MANTA TEST SYSTEMS
CU A002 15A
FIGURE 4.30 EXTERNAL AMP SIGNAL INPUT CONNECTOR
Pin #
Signal
1
Va input
2
Signal Ground
3
Vb input
4
Vc input
5
Ia input
6
External Sync +
7
External Sync -
4.11
AUDIBLE TONE
The audible tone is used for multiple purposes:
1. To indicate relay operation
This function can be enabled/disabled by pressing the [TONE] button.
It’s enabled when a long beep sounds after pressing the [TONE] button. It’s disabled when a short beep
sounds after pressing the [TONE] button.
When enabled, and when a closed contact or a voltage (above 10V) is sensed on the appropriate
[EXTERNAL STOP] trigger inputs, the tone will be activated. The tone will persist until the triggering
source is removed.
This feature is commonly used in static operation mode for manual pickup checks.
2. To indicate entry into the postfault state
As with the previous function, this function can be enabled/disabled by pressing the [TONE] button.
When enabled, and when the MTS-1710 enters the postfault state, a short beep occurs to indicate completion
of a dynamic test. Of course, this only occurs in dynamic operation mode.
Pin 6
Pin 1
Pin 4
Pin 2
Pin 7
Pin 3
Pin 5
External Amp Signal Input Connector
As viewed from rear panel of the instrument
Содержание MTS-1700 SERIES
Страница 30: ...TABLE OF CONTENTS xxx MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 38: ...INTRODUCTION Section 1 1 8 MTS 1700SERIESOPERATIONANDREFERENCEMANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 44: ...SPECIFICATIONS Section 2 2 6 MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 248: ...FAULT PLAYBACK Section 7 7 12 MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 262: ...SERVICING Section 8 8 14 MTS 1700 SERIES OPERATION ANDREFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 320: ...PROGRAMMABLE DIGITAL I O CHANNELS A 14 MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 340: ...E 4 MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 344: ...GLOSSARY F 4 MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...