![Manta MTS-1700 SERIES Скачать руководство пользователя страница 139](http://html1.mh-extra.com/html/manta/mts-1700-series/mts-1700-series_operation-and-reference-manual_1723616139.webp)
DETAILED OPERATION - Section 4
MTS-1700 SERIES OPERATION AND REFERENCE MANUAL 4-49
CU A002 15A
MANTA TEST SYSTEMS
4.14
DC VOLTAGE CONTROL
This section only applies to units with the DC Voltage output option.
DC voltage is available at the [DC VOLTS] terminals of the front panel. This output may be live (depending
upon the setting) whenever the MTS-1710 is on since it’s intended to supply power to solid state relays
under test. A maximum of 100W and 300V is available.
Displaying the DC voltage level:
Press the [VOLTAGE] button once or twice (as required) to activate the DC voltage display.
The units annunciator should read “DCVolts” in the [MODE/MENU DISPLAY], as illustrated below:
48.0
DCVolts IST LINE
B-N
Φ
-N I3
Press the [VOLTAGE] button again to return to the AC voltage display. To prevent accidental adjustment
of the DC voltage, adjustments can only be made through the menu.
Enabling the output (Enable button/LED indicator):
For units with the additional output enable button and LED indicator, the button to the left of the output
terminals enables and disables the output.
The LED colour indicates the following ranges:
LED Colour Output
OFF OFF
Green 24-70V
Yellow 70-130V
Red 130-300V
Setting standard DC voltage output levels:
Activate the menu using the [MODE/MENU] button, and select SETTINGS DC-VOLTS in the menu. The
[MODE/MENU DISPLAY] will appear as shown below.
Turn the [MODIFY] knob to position the cursor over one of the OFF, 24V, 48V, 125V or 250V selections.
Press [SELECT] and the selected level will be set, and the present measured output will be displayed.
Fine adjustment of the level can be performed by turning the [MODIFY] knob. Press [MODE/MENU] when
done.
Содержание MTS-1700 SERIES
Страница 30: ...TABLE OF CONTENTS xxx MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 38: ...INTRODUCTION Section 1 1 8 MTS 1700SERIESOPERATIONANDREFERENCEMANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 44: ...SPECIFICATIONS Section 2 2 6 MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 248: ...FAULT PLAYBACK Section 7 7 12 MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 262: ...SERVICING Section 8 8 14 MTS 1700 SERIES OPERATION ANDREFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 320: ...PROGRAMMABLE DIGITAL I O CHANNELS A 14 MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 340: ...E 4 MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 344: ...GLOSSARY F 4 MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...