RS-232C INTERFACE - Section 6
6-4 MTS-1700 SERIES OPERATION AND REFERENCE MANUAL
MANTA TEST SYSTEMS
CU A002 15A
RES
Reset
•
Returns the MTS-1710 to prefault state.
•
Clears any overload warnings which may exist.
PRF#
Set Prefault on/off
•
Valid values are 0 and 1
0 - sets prefault V&I off
1 - sets prefault V&I on
POF#
Set Postfault on/off
•
Valid values are 0 and 1
0 - sets postfault V&I off
1 - sets postfault V&I on
PFS
Print Fault State
•
The MTS-1710 will return the current fault state (“PREFAULT”, “FAULT” or
“POSTFAULT”)
•
Example:
Ready>DYN
<- select dynamic operation mode
Ready>PFS
<- interrogate fault state
PREFAULT
Ready>STR
<- cause start trigger
Ready>PFS <- interrogate fault state
FAULT
Ready>PTS <- print timer reading
2.104sec
Ready>STP <- cause stop trigger
Ready>PFS <- interrogate fault state
POSTFAULT
Ready>
STS
Print Stop Trigger Status
•
Returns the status of the stop trigger input.This is used to sense the presence of
a closed contact or voltage on the [EXTERNAL STOP] trigger inputs.
•
If a stop trigger channel is selected on the digital I/O channels, the selected
channel is also checked for closed contact or voltage.
•
“ACTIVE” is returned if a closed contact or voltage is sensed.
•
“INACTIVE” is returned if neither of the above is sensed.
WFS#
Wait on Fault State
•
Wait for specified time or exit from FAULT state.
•
Waits until either the fault state changes to prefault or postfault, or until the
specified time elapses. After this, the present fault state is returned.
•
Valid values are 0.02 to 99.99 seconds.
•
During the waiting period, RS-232 command execution is suspended.
•
This command is meant to be executed only in FAULT state, with operation
mode set to dynamic. The delay value is programmed with the maximum
expected relay operation time.
Содержание MTS-1700 SERIES
Страница 30: ...TABLE OF CONTENTS xxx MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 38: ...INTRODUCTION Section 1 1 8 MTS 1700SERIESOPERATIONANDREFERENCEMANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 44: ...SPECIFICATIONS Section 2 2 6 MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 248: ...FAULT PLAYBACK Section 7 7 12 MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 262: ...SERVICING Section 8 8 14 MTS 1700 SERIES OPERATION ANDREFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 320: ...PROGRAMMABLE DIGITAL I O CHANNELS A 14 MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 340: ...E 4 MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 344: ...GLOSSARY F 4 MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...