PROGRAMMABLE DIGITAL I/O CHANNELS
MTS-1700 SERIES OPERATION AND REFERENCE MANUAL A-11
CU A002 15A
MANTA TEST SYSTEMS
Example 3 Multiple channel digital output sequence.
The following command sequence programs output channels 0-3 to output pre-defined sequence, as
illustrated in Figure A-1(c).
DIO,OCL
<- clear output settings
DYN
DIO,OUT0,0,HF
<- channels 0-3 prefault level=0
DIO,OUT1,1,1
<- channel 1 changes to 1 on FAULT
DIO,ODY25,%1010,%1011
<- channels 0,1 & 3 change at 25ms
DIO,ODY40,%100,%100
<- channel 2 changes to 1 at 40 ms
DIO,ODY42,0,%10
<- channel 1 changes to 0 at 42 ms
DIO,ODY106,0,%1000
<- channel 3 changes to 0 at 106ms
DIO,OUT2,0,HF
<- channels 0-3 postfault level=0
STR
<- initiate fault using STR command (or use external trigger)
Example 4 Simulated Breaker Closed (52A) Signal.
The following command sequence sets channel 0 to be “off” in postfault state and “on” in all other states.
To simulate a 52A signal:
DIO,OCL
<- clear output settings
DIO,OUT0,1,1
<- set channel 0 to be on in prefault
DIO,OUT2,0,1
<- set channel 0 to be off in postfault
Note that no command is required to set the fault state value since no change is required when the fault state
changes from prefault to fault.
Example 5 Simulated Breaker Open (52B) Signal.
The following command sequence sets channel 0 to be “on” in postfault state, and “off” in all other states.
To simulate a 52B signal:
DIO,OCL
<- clear output settings
DIO,OUT0,0,1
<- set channel 0 to be on in prefault
DIO,OUT2,1,1
<- set channel 0 to be off in postfault
Note that no command is required to set the fault state value since no change is required when the fault state
changes from prefault to fault.
A.6.6 Phase Comparison Output Programming
A.6.6.1 COMMANDS.
Command
Description
DIO,PCO#
Phase comparison output enable/disable
•
Valid values are 0-1
•
0 = disabled (AUXOUT15 becomes normal digital output)
•
1 = enabled (AUXOUT15 becomes programmable square waveoutput)
•
The default value is 0 (disabled)
Содержание MTS-1700 SERIES
Страница 30: ...TABLE OF CONTENTS xxx MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 38: ...INTRODUCTION Section 1 1 8 MTS 1700SERIESOPERATIONANDREFERENCEMANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 44: ...SPECIFICATIONS Section 2 2 6 MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 248: ...FAULT PLAYBACK Section 7 7 12 MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 262: ...SERVICING Section 8 8 14 MTS 1700 SERIES OPERATION ANDREFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 320: ...PROGRAMMABLE DIGITAL I O CHANNELS A 14 MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 340: ...E 4 MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 344: ...GLOSSARY F 4 MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...