![Manta MTS-1700 SERIES Скачать руководство пользователя страница 100](http://html1.mh-extra.com/html/manta/mts-1700-series/mts-1700-series_operation-and-reference-manual_1723616100.webp)
DETAILED OPERATION - Section 4
4-10
MTS-1700 SERIES OPERATION AND REFERENCE MANUAL
MANTA TEST SYSTEMS
CU A002 15A
4.5.7
Two-Wire Pulse Timing
By parallelling the START and STOP inputs, pulse type operations may be timed using only a single pair
of sensing leads.
The rising edge of a voltage pulse, for example, would cause a start trigger, and the falling edge would cause
a stop trigger. This allows measurement of the duration of a voltage pulse.
FIGURE 4.8 TWO-WIRE PULSE TIMING CONNECTIONS
4.5.8
Timing in Cycles
The [TIME] button toggles between timing in seconds and timing in cycles display.
Timing in cycles is calculated from the timer (seconds) reading and the present frequency. In line frequency
mode, there must be a voltage output to obtain a frequency measurement, and subsequently a timing in
cycles value.
If the frequency changes during a test, the time in cycles value will assume that the frequency has always
stayed at the latest value (time in cycles doesn’t actually count power cycles--it’s a calculated value).
VOLTAGE PULSE INPUT
OR
CONTACT OPEN-CLOSE-OPEN
OR CLOSE-OPEN-CLOSE INPUT
Time
PARALLELED START & STOP INPUTS FOR TWO-WIRE TIMING
V
Select EXTERNAL START timing mode
via the front panel menu.
NOTE:
Содержание MTS-1700 SERIES
Страница 30: ...TABLE OF CONTENTS xxx MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 38: ...INTRODUCTION Section 1 1 8 MTS 1700SERIESOPERATIONANDREFERENCEMANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 44: ...SPECIFICATIONS Section 2 2 6 MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 248: ...FAULT PLAYBACK Section 7 7 12 MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 262: ...SERVICING Section 8 8 14 MTS 1700 SERIES OPERATION ANDREFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 320: ...PROGRAMMABLE DIGITAL I O CHANNELS A 14 MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 340: ...E 4 MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...
Страница 344: ...GLOSSARY F 4 MTS 1700 SERIES OPERATION AND REFERENCE MANUAL MANTA TEST SYSTEMS CU A002 15A ...