Ch.3 Settings
< 59 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
3.5
DICOM SCU Settings
The procedure for setting DICOM SCU (DICOM information for diagnostic devices connected to
the DRYPRO 771) using the service maintenance mode, is described below.
The following DICOM SCU items are available for setting.
• S11 CHANNEL USE
: This setting determines whether or not to use the channel of the selected
diagnostic device.
"ON" or "OFF" may be selected.
default setting : OFF
• S12 SCU NAME
: Sets the name of the selected diagnostic device.
A maximum of 12 alphanumerical characters may be input.
The name set here is used in the DRYPRO 771 only and will be displayed as
the diagnostic device name.
• S13 N-EVENT PORT
: Sets the N-EVENT PORT number.
default setting : 00000
• S14 IP ADDRESS
: Sets the IP address of the selected diagnostic device.
default setting : 0.0.0.0
• S15 AE TITLE
: Sets the AE title of the selected diagnostic device.
A maximum of 16 alphanumerical characters may be input.
default setting : blank
• S16 HOST NAME
: Sets the name of the selected diagnostic device to be used on the network.A
maximum of 16 alphanumerical characters may be input.
default setting : blank
• S17 BORDER SIZE
: Sets the output film border size."REGIUS" (no margin) or "CT/MRI" (with
margin) may be selected.
default setting : CT/MRI
"REGIUS"
: Border with almost no margin used, for example, with
REGIUS CR.
"CT/MRI"
: Sets a margin suitable for CT or MR multi-format images.
• S18 FILM SIZE/TYPE
: This setting determines whether or not a notification of film size and type check
is given.
"NO CHECK" or "CHECK" may be selected.
default setting : CHECK
• S19 EMPTY NOTICE
: This setting determines whether or not notification is sent to the diagnostic
device when the supply tray is empty.
"ON" or "OFF" may be selected.
default setting : ON
• S1A DICOM ERR LV
: Sets the method of return of DICOM errors from the DRYPRO 771 to diagnostic
devices.
"STANDARD," "FAILURE" or "N-ACTION" may be selected.
default setting : STANDARD
"STANDARD"
: When the diagnostic device checks the DRYPRO 771 status
(N-GET), the DRYPRO 771 returns to the diagnostic device a
"WARNING" signal to indicate, for example, that the supply
tray is empty and a "FAILURE" signal for serious errors.
"FAILURE"
: A "FAILURE" signal is returned for all errors.
"N-ACTION"
: In cases where the diagnostic device does not check the
DRYPRO 771 status (N-GET), the DICOM connection is
automatically broken (association aborted) by a print
command (N-ACTION).
S10 DICOM SCU /.
Service Manual
Web Maintenance
Tool>
SCU>
DICOMSCU
Содержание DRYPRO 771
Страница 10: ...Preface 8 DRYPRO MODEL 771 INSTALLATION MANUAL Ver 1 00 2003 10...
Страница 26: ...Ch 1 Before Installation 24 DRYPRO MODEL 771 INSTALLATION MANUAL Ver 1 00 2003 10...
Страница 54: ...Ch 2 Unpacking and Installation 52 DRYPRO MODEL 771 INSTALLATION MANUAL Ver 1 00 2003 10...
Страница 82: ...Ch 4 Checking the Image 80 DRYPRO MODEL 771 INSTALLATION MANUAL Ver 1 00 2003 10...
Страница 96: ...Ch 4 Checking the Image 94 DRYPRO MODEL 771 INSTALLATION MANUAL Ver 1 00 2003 10...