Ch.3 Settings
< 76 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
S70 SYSTEM SETUP /.
p.55 Starting the
Service Maintenance
Mode
U80 SERVICE MODE /.
S00 DICOM SCP /.
.
.
.
.
S71 PATCH CONTROL /.
OFF
S40 FILM SETUP /.
1
Start the maintenance mode.
2
Display "U80 SERVICE MODE", and press the
[enter] key.
•
"U00 DICOM SCP" will be displayed on the
message display window.
3
Press [
d
] of the [menu/select] key seven times.
•
"U70 SYSTEM SETUP" will be displayed.
4
Press the [enter] key.
•
Input screen will be displayed.
5
Input the value or text according to the input box
of each item.
•
Use the [menu/select] key
[
d
][
c
]
to choose item to
be input.
•
"S71 PATCH CONTROL" screen is an example for
density control ON or OFF.
6
Press the [enter] key after completing the input.
•
Settings will be registered and display returned to
the maintenance menu screen.
When [exit] key is pressed in the "SYSTEM SETUP", a confirmation screen will be
displayed. Select "YES" or "NO" for the register of the changed data using [select]
key [
[
][
\
].
When "YES" selected
: Changes to the setting (data) are registered, and the
display returns to the maintenance menu screen.
When "NO" selected
: Display returns to the maintenance menu without
registering the changes to the setting (data).
Data was changed
Save the data? YES
Содержание DRYPRO 771
Страница 10: ...Preface 8 DRYPRO MODEL 771 INSTALLATION MANUAL Ver 1 00 2003 10...
Страница 26: ...Ch 1 Before Installation 24 DRYPRO MODEL 771 INSTALLATION MANUAL Ver 1 00 2003 10...
Страница 54: ...Ch 2 Unpacking and Installation 52 DRYPRO MODEL 771 INSTALLATION MANUAL Ver 1 00 2003 10...
Страница 82: ...Ch 4 Checking the Image 80 DRYPRO MODEL 771 INSTALLATION MANUAL Ver 1 00 2003 10...
Страница 96: ...Ch 4 Checking the Image 94 DRYPRO MODEL 771 INSTALLATION MANUAL Ver 1 00 2003 10...