Specification
< 97 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
When Connected to REGIUS
When Connected to Units Other than REGIUS
14 X 17 inch
4,503 x 5,419 pixel
4,351 x 5,114 pixel
14 X 14 inch
4,500 x 4,452 pixel
4,351 x 4,236 pixel
11 X 14 inch
3,498 x 4,503 pixel
3,396 x 4,236 pixel
Image Matrix
Number of
1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 25, 30, 35, 36, 42, 48, 54, 56, 60, 63, 64
Recorded Frames
Positive/Negative
Available.
Trim
Available.
Border
Black or Clear.
Stamp
Recorded on 1 line or 2 line of film (May be printed at top/bottom/left/right).
Density
Furnished in main body.
Correction
Start timer
Day/time settings possible.
Applicable
IEC60601-1-2 : 2001, IEC60601-1, IEC60825
Standards
Remarks
DRYPRO 771 is manufactured in factories certified in compliance with ISO9001 : 2000, EN
ISO13485 : 2000, 93/42/EEC quality control standards as well as with medical device directive
FDA Pre market Notification 510(k) and GMP.
Accessories
Deodorant filter unit, Suction cups (spare), cutter, One-touch Spacers (2 pcs), Power cable,
Clamp (2 pcs), Exhaust duct, Warranty, Operation manual, Operation sheet, Attached
document for doctors, Inspection sheet, Protective parts(red) collection bag,
Accessory list
Options
Deodorant filter
(available at cost)
Warranty
12 months after installation.
Содержание DRYPRO 771
Страница 10: ...Preface 8 DRYPRO MODEL 771 INSTALLATION MANUAL Ver 1 00 2003 10...
Страница 26: ...Ch 1 Before Installation 24 DRYPRO MODEL 771 INSTALLATION MANUAL Ver 1 00 2003 10...
Страница 54: ...Ch 2 Unpacking and Installation 52 DRYPRO MODEL 771 INSTALLATION MANUAL Ver 1 00 2003 10...
Страница 82: ...Ch 4 Checking the Image 80 DRYPRO MODEL 771 INSTALLATION MANUAL Ver 1 00 2003 10...
Страница 96: ...Ch 4 Checking the Image 94 DRYPRO MODEL 771 INSTALLATION MANUAL Ver 1 00 2003 10...