Ch.3 Settings
< 60 >
DRYPRO MODEL 771 INSTALLATION MANUAL Ver.1.00 2003.10
• S1B FRAME LAYOUT
: Sets the spacing between frames in multi-format output.
"EVEN"(even layout) "CENTER" (centered layout), "H EVEN" (even layout, Hitachi
format) or "H CENTER" (centered layout, Hitachi format) may be selected.
default setting : EVEN
"EVEN"
: Film is printed with the top/bottom/left/right margins unchanged
and the spaces between frames as wide as possible. (Even
layout)
"CENTER"
: Film is printed with the spaces between frames uniform and the
top/bottom/left/right margins broadened. (Centered layout)
"H EVEN"
: Selected to produce even layout when a Hitachi diagnostic
device is connected.
"H CENTER"
: Selected to produce centered layout when a Hitachi diagnostic
device is connected.
• S1C DMAX/DMIN
: This setting determines whether maximum and minimum densities will be printed
according to the DRYPRO 771 settings or the diagnostic device settings.
"DICOM" (diagnostic device) or "P-CON" (DRYPRO 771) may be selected.
default setting : DICOM
• S1D P-VALUE RCV
: Sets the tone conversion method of the image.
Select from "OFF" (as in the past) and "ON" (tone conversion using "P" value) .
default setting : OFF
Relation among Status of the DRYPRO 771, "DICOM ERR LV" setting and communication with external
devices.
DICOM ERR LV
DICOM Comm. Message
Queue Reg.
Status of DRYPRO 771
displayed on external device
STANDARD
NORMAL
A/V
READY
FAILURE
NORMAL
A/V
READY
N-ACTION
NORMAL
A/V
READY
Ready for print • warming up
(
including the case where "S19 EMPTY NOTICE" is set to "OFF" and film empty.
DICOM ERR LV
DICOM Comm. Message
Queue Reg.
Status of DRYPRO 771
displayed on external device
STANDARD
FAILURE
N/A
ERROR
FAILURE
FAILURE
N/A
ERROR
N-ACTION
NORMAL
A/V
READY
In User or Service Maintenance Mode
DICOM ERR LV
DICOM Comm. Message
Queue Reg.
Status of DRYPRO 771
displayed on external device
STANDARD
WARNING
note 2
WARNING
FAILURE
FAILURE
N/A
ERROR
N-ACTION
NORMAL
note 3
READY
Film Empty (S19 EMPTY NOTICE is set to "ON")
DICOM ERR LV
DICOM Comm. Message
Queue Reg.
Status of DRYPRO 771
displayed on external device
STANDARD
FAILURE
N/A
ERROR
FAILURE
FAILURE
N/A
ERROR
N-ACTION
NORMAL
note 3
READY
In Error Status
note 2
: Queue registration is available when output under "WARNING" is set to be allowed on the external diagnostic device.
note 3
: Forcefully interrupt the DICOM connection using the print command (N-ACTION).
Содержание DRYPRO 771
Страница 10: ...Preface 8 DRYPRO MODEL 771 INSTALLATION MANUAL Ver 1 00 2003 10...
Страница 26: ...Ch 1 Before Installation 24 DRYPRO MODEL 771 INSTALLATION MANUAL Ver 1 00 2003 10...
Страница 54: ...Ch 2 Unpacking and Installation 52 DRYPRO MODEL 771 INSTALLATION MANUAL Ver 1 00 2003 10...
Страница 82: ...Ch 4 Checking the Image 80 DRYPRO MODEL 771 INSTALLATION MANUAL Ver 1 00 2003 10...