Status Reporting
88
Table 8-2. Bit Configurations of Status Registers
Bit
Signal
Meaning
Bit
Signal
Meaning
Operation Status Group
Standard Event Status Group
0
CAL
The interface is computing new
calibration constants.
0
OPC
Operation complete.
5
WTG
The interface is waiting for a trigger.
2
QYE
Query error.
8
CV
The power module is in constant
voltage mode.
3
DDE
Device-dependent error.
10
CC
The power module is in constant
4
EXE
Execution error.
current mode.
5
CME
Command error.
7
PON
Power on.
Questionable Status Group
Status Byte and Service Request
Enable Registers
0
OV
The power module overvoltage
protection circuit has tripped.
3
QUES
Questionable status summary bit.
1
OC
The power module overcurrent
protection circuit has tripped.
4
MAV
Message Available summary bit.
4
OT
The power module has an
overtemperature condition.
5
ESB
Event Status summary bit.
9
RI
The power module remote inhibit state
is active.
6
MSS
RQS
Master Status summary bit.
Request Service bit.
10
UNR
The power module output is
unregulated.
7
OPER
Operation status summary bit.
Figure 8-1. Power supply Status Model