Language Dictionary 73
Current Subsystem
This subsystem programs the output current of the power supply.
CURR
CURR:TRIG
These commands set the immediate current level or the pending triggered current level of the power supply. The immediate
level is the current programmed for the output terminals. The pending triggered level is a stored current value that is
transferred to the output terminals when a trigger occurs. A pending triggered level is unaffected by subsequent
CURR
commands and remains in effect until the trigger subsystem receives a trigger or an
ABORt
command is given. If there is
no pending
CURR:TRIG
level, then the query form returns the
CURR
level. In order for
CURR:TRIG
to be executed,
the trigger subsystem must be initiated (see
INITiate).
Command Syntax
[SOURce]:CURRent[:LEVel] [:IMMediate][:AMPLitude] <NRf+>
[SOURce][:CURRent[:LEVel]:TRIGgered [:AMPLitude] <NRf+>
Parameters
Table 7-1
Default Suffix
A
*RST Value
Table 7-1
Examples
CURR 200
MA
CURRENT:LEVEL 200 MA
CURRENT:LEVEL:IMMEDIATE:AMPLITUDE 2.5
CURR:TRIG 20
CURRENT:LEVEL:TRIGGERED 20
Query Syntax
[SOURce]:CURRent[:LEVel] [:IMMediate][:AMPLitude]?
[SOURce]:CURRent[:LEVel] [:IMMediate][:AMPLitude]? MAX
[SOURce]:CURRent[:LEVel] [:IMMediate][:AMPLitude]? MIN
[SOURce]:CURRent[LEVel]:TRIGgered [:AMPLitude]?
[SOURce]:CURRent[LEVel]:TRIGgered [:AMPLitude]? MAX
[SOURce]:CURRent[:LEVel]:TRIGgered [:AMPLitude]? MIN
Returned Parameters
<NR3> CURR?
and
CURR:TRIG?
return presently programmed immediate and
triggered levels. If not triggered level is programmed, both returned values are the same.
CURR? MAX
and
CURR? MIN
return the maximum and minimum programmable
immediate current levels.
CURR:TRIG? MAX
and
CURR:TRIG? MIN
return the maximum and minimum
programmable triggered current levels.
Related Commands
For
CURR
*SAV *RCL
*RST
For
CURR:TRIG
ABOR CURR *RST
CURR:PROT:STAT
This command enables or disables the power supply overcurrent protection (OCP) function. If the overcurrent protection
function is enabled and the power supply goes into constant-current operation, then the output is disabled and the
Questionable Condition status register OC bit is set (see "Chapter 8 - Status Reporting"). An overcurrent condition can be
cleared with the
OUTP:PROT:CLE
command after the cause of the condition is removed.
Command Syntax
[SOURce]:CURRent:PROTection:STATe <bool>
Parameters
0 | 1 | OFF | ON
*RST
Value
OFF
Examples
CURR:PROT:STAT 0 CURRENT:PROTECTION:STATE OFF
CURR:PROT:STAT 1 CURRENT:PROTECTION:STATE ON
Query Syntax
[SOURce]:CURRent:PROTection:STATe?
Returned Parameters
<NRI>
0
or
1
Related Commands
OUTP:PROT:CLE *RST